Whole Blood, Serum, Plasma or UrineSample Size100 l Whole Blood, Serum or PlasmaNumber of Parameters3 SimultaneouslyReproducibilityTypical Within Run n30Blood, Serum, PlasmaNa CV 1 115-160mmol/LK CV 2 2.5-6 mmol/LCl- CV 2 65-120 mmol/LUrineNa CV 5 60-280 mmol/LK CV 5 30-170 mmol/LCl- CV 5 80-230 mmol/LLi SD 0.03 0.25-2.0 mmol/LDisplay ResolutionNa 1 mmol/LK 0.1 mmol/LCl- 1 mmol/LLi 0.01 mmol/LAnalysis Time57 sec Blood93 sec UrineData Storage1000 Patient Results500 Level 1 QC Results500 Level 2 QC Results500 Level 3 QC ResultsCalibration2 Point every 4 hoursOn DemandUser InputKeypad, External KeyboardExternal Barcode Scanner optionalOutputGraphic backlit displayGraphic 40 column printerRS232 LIS CommunicationDetection Range 45-205 mmol/L 1.5-11 mmol/L 45-205 mmol/L 0.15-5.0 mmol/L 30-1020 mmol/L 20-505 mmol/LBlood Na KClLiUrine NaKCl- 25-505 mmol/L.
Other Specifications Download Specification | |
---|---|
System Is Capable To Measure Alkali And Alkaline Earth Metals (Sodium, Potassium, Calcium, Lithium) In The Samples | Yes |
Ignition System | Auto ignition |
Analyzed Elements | Na+, Ca++, K+, Li+ |
Accuracy | Less than +- 2 % |
Reproducibility | Less than +- 2 % |
Sensitivity | Na+ = 1 ppm, Ca++ = 15 ppm, K+ = 1 ppm, Li+ = 1 ppm |
Display | at least 3 lines LCD |
Detector | Photo conductive cell |
Filters | Sodium, Potassium, Calcium, Lithium |
Aspiration Rate | 3 to 6 ml/min |
Memory | To store and recall 800 samples |
Type Of Printer Interface For Printing Of Samples | Thermal |
Accessories | Compressor with built in air regulator and air filter |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | Yes |
Power Requirements | 230 V +- 10, 50 Hz AC |
Operating Temperature | 15-32 Degree C |
Operating Humidity (RH) | < 85 % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | yes |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | yes |
Test Report No. | 02014 |
Test Report Date | 24.08.2017 |
Name Of The Lab | Central Lab |
Warranty | 1 Year |