Autoclave will be in sturdy construction using plates of appropriate thickness which prevents buckling under vacuum or bulging due to pressure. Made to confirm to EN 285 or HTM 2010 performance requirements. Fully automatic by means of PLC based control. Autoclave for HPHV cycle as per General Technical Specification. Door Movement vertical, Smooth Noiseless Door Operating. No Human Effort required to the Door. Chamber Dimension Is 610x910x1515 mm and chamber capacity is 840 Liter. On line printing recording confirming to 21 CFR PART 11. Options like electrical steam generator, vacuum pump, carriage, trolleys, chart recorders, data loggers, Dx station. Heat up to Range 121 to 134 Degree C. It confirmed standard IS 13489, ICE 60601-1 and ICE 60601-2..
Other Specifications Download Specification | |
---|---|
Sterilizer, Shallow (Dressing Drum) Conforming To IS: 3831-1979 | yes |
Weather ISI Marked | No |
If Yes, CM/L No. | 0 |
Material Of Body, Lid & Belt And Hasp | Stainless steel |
Designation Of Stainless Steel | 04Ce18Ni10 to IS:6911-1992 |
Finish Of Stainless Steel | 2B as per IS:6911-1972 |
Shape | Cylirical |
Height Of Sterilizer Mm | 350 mm |
Diameter Of Sterilizer Mm | 240 mm |
Avaliability Of Test Report Fron Govt./ NABL / ILAC Laboratory To Prove Conforming To Specification. | yes |
Test Report No. And Date | 2074/C0482 |