"Product Catgeory","signal generator" "Product Name","Keysight N5173B-540 RF Signal Generator 40 GHz" "Specification Title","Specification Value" Type,"RF Signal Generator" "Frequency Range","10 MHz to 40 GHz" "Frequency Resolution","0,1" "Ageing Rate for reference frequency (As declared by OEM)","Less than 1" "Temperature stability for reference frequency","Less than 1" "SSB Phase Noise at 1GHz,on 20KHz offset (dBc/Hz)","less than -105" "Harmonics at 3 GHz level at 0 dBm","less than -30" "Non Harmonics (>10 KHz carrier offset)","less than -50" "Non Harmonics (>50 KHz carrier offset)",NA "Reverse Power Protection","0.5 W, 0 Vdc Watt" "Minimum Output Level",-120 "Maximum Output Level",+5 "Output Level Resolution","0,1" "Total Accuracy forall specified Output Level at 25 Degree C in continuous wave mode","± 0.9 dB dB" "VSWR on all operating frequencies",1.6 "Memory (Minimum Storage Settings)",1000 "Amplitude Modulation (AM) Operating Modes","Internal & External" "Minimum Amplitude Modulation Frequency (3 dB), f > 100 KHz","DC or 10 Hz Hz" "Maximum Amplitude Modulation Frequency (3 dB), f > 100 KHz","100 kHz" "Modulation Depth","0 to 100%" "Resolution of AM",0.1% "AM Distortion at 1 KHz","< 1.5% of setting +1%" "Frequency Modulation (FM) Operating Modes","Internal & External" "Minimum Frequency Modulation Frequency (-3 dB) Standard","DC or 1 Hz" "Maximum Frequency Modulation Frequency (-3 dB) Standard","7 MHz" "Minimum Frequency Deviation",2.5MHz "Maximum Frequency Deviation",320MHz "Resolution of FM","0.025% of deviation or 1 Hz" "Setting Error (at AF=1 KHz and 50% max. Deviation) for FM","< ± 2% + 20 Hz" "Phase Modulation",Available "If Available, Phase Modulation Operatinf Mode","Internal & External" "Minimum Phase Deviation","1.25 Radians Radians" "Maximum Phase Deviation","160 Radians Radians" "Resolution of Phase Modulation",1 "Setting Error (at AF=1 KHz) for Phase Modulation","< 0.2% (typ)" "Internal Modulation Generator For AM/FM/PM","0.1 Hz to 2 MHz tunable to 3 MHz with resolution of 0.1 Hz" "Pulse Modulation",Not-Available "If Available, Pulse Modulation Operating Modes",NA "Rise and Fall Time",NA "On/Off Ratio",NA "Minimum Pulse Width",NA "Input Level",NA "RF Digital Sweep Range and Width","9 KHz to 40GHz" "RF Digital Sweep Time Range","100 μs to 100s" "RF Analog sweep",Not-Available "If Available, RF Analog Sweep Range and Width",NA "RF Analog Sweep Time Range",NA "Pulse Generator",Not-Available "If Available, Pulse Generator Operating Mode",NA "Pulse Width",NA "Pulse Delay",NA "Pulse Period",NA "Vector Modulation",Available "Internal Base Band Generator Bandwidth",NA "Arbitory waveform memory",NA Modulation,NA "Error Vector Magnitude (EVM)",0 "Phase Modulator",Available "If available, Range of Modulating Signal (Wide Mode)",NA "If available, Range of Modulating Signal (Narrow Mode)",NA "Interface Available","GPIB IEEE-488.2, 1987 with listen and talk LAN 1000BaseT LAN interface, LXI Class C compliant USB Version 2.0" "Standard Acessories (To be supplied)","Country-specific power cord Documentation set (CD-ROM) 3-year return-to-Keysight warranty" "Power Requirements","230 V +/-10% AC, 50 Hz +/- 3%" "Operating Temperature Range","0 to 55 Degree C Degree C" "Operating Humidity (RH)","Yes Available" ""Availability of Test Reports from Central Govt / NABL approved / ILAC acredited lab to prove conformity to the specification including environmental tests as under : 1. Dry Heat Test : 45 deg C for 16 hours as per IS:9000/Pt.3/sec.5/1977 (reaffirmed in 2007) 2. Cold test : 0 deg C for 4 hours as per IS:9000/Pt.2/sec.4/1977 (reaffirmed in 2007) 3. Damp Heat Test: 95% RH at 45 deg C, two cycles of 24 hours each separated by a gap of 4 hours as per IS:9000/Pt.5/sec. 1/1991 (reaffirmed in 2007) All the functional paramteres are to be verified before and after recovery period of 1 hours of above environmental test sequence as above. There shall be no deterioration in functional parameters.",Yes "Test Report No.",4130591-4927981-1 "Test Report Date",17-JAN-2015 "Name of the Lab","Keysight Technologies Malaysia" Warranty,3