× All the transactions performed on 2.0 is available here. To access Market click "Home".

cryo bath (mlabs)(mlabs-cb)

` 345,999.00

  • MRP: `416500.00

Temperature range from -5C to 100C with micro controller based temperature controller includes digital temperature display of set value and process value. 1.The cabinet of N and M cryobath is insulated with high density CFC free Poly Urethane Foam 120 mm thick insulation.2.Interior chamber is finished resistant stainless Steel 304, 0.8 mm thick.3.The exterior is Sky Blue/white Pure Polyester powder coated 1.0 mm thick Mild Steel and the door has a magnetic gasket with a keyed lock.4.Acrylic trays are provided for holding plasma bags.5. High density CFC Free Poly Urethane Foam insulated cabinet with corrosion resistant stainless steel interior and white powder coated exterior finish. Temp Controller Micro processor based temperature control with accuracy of 1C. ISO 9001, ISO1400, ISO13485, CE and WHO-GMP certified..

  • Availability: 1,000 In Stock
  • Min. Qty. Per Consignee: 1
  • Minimum numbers of thermometers which can be simultaniously calibrated in water bath. - 20 Nos
  • Reference thermometer confirming to IS:4825/1982 - YES
  • Min. resolution of reference thermometer - 0.01 °C
  • The Comparator water bath either screw type or bubble type as per IS:6274/1971,should be constant and uniform within ±0.01 °C,stability ±0.01 °C and should have digital indication with resolution of 0.01 °C. Thermocouple shall be calibrated and shall have calibration certificate. - Yes
  • Centrifuge to drive mercury level well below 25C at 600 rpm. - Yes
Other Specifications Download Specification
Purpose Of Waterbath. To test clinical thermometer (Liquid in glass(LIG) &electrical) with maximum device within accuracy of 0.15 C
Minimum Numbers Of Thermometers Which Can Be Simultaniously Calibrated In Water Bath. 20 Nos
Reference Thermometer Confirming To IS:4825/1982 YES
Min. Resolution Of Reference Thermometer 0.01 °C
Min.uncertainty Of Reference Thermometer(at Coverage Factor K=2) 0.02 °C
The Comparator Water Bath Either Screw Type Or Bubble Type As Per IS:6274/1971,should Be Constant And Uniform Within ±0.01 °C,stability ±0.01 °C And Should Have Digital Indication With Resolution Of 0.01 °C. Thermocouple Shall Be Calibrated And Shall Have Calibration Certificate. Yes
Mechanism To View The LIG Thermometer With Magnification Of 10X. Yes
Mechanism To Hold Thermometer In Water For At Least 20 Seconds. Yes
Centrifuge To Drive Mercury Level Well Below 25C At 600 Rpm. Yes
Thermometer Calibrator Shall Be Capable To Calibrate The LIG Thermometer As Per IS:12622/1989 Yes
All The Gazette Of India Related To Legal Metrology Issued Time To Time Are Applicable To The Specification As Well As To The Product.Also As Per The Gazette Of India Extraodinary,The Legal Metrology Act 2009,cl.22, Approval Of Model Before Manufacturing And Cl.23.Prohibition On Manufacture,repair Or Sale Without Licence For Product Shall Be Applicable On Manufacture Of Measuring Instrument. Yes
Availability Of Calibration Test Reports Of The Reference Thermometer From NPL Or Central Govt/NABL/ILAC Accredited Lab . Yes
Availability Of Type Test Reports Of The Product From Central Govt/NABL/ILAC Accredited Lab Showing Compliance To The Specification Yes
Test Reports No. 42315
Test Reports Date 21/05/2017
Name Of The Labs City Lab Centre
Address Of The Lab Dehradun
Product Code Available On Supplier Website Yes
Product Code No mLabs-CB

Price Dekho Price Dekho Price List in india