Pathological TRinocular Microscope ISI Marked To IS 4381/1967 8275/1976Reaffirmed 2012 With Amdts, 1 2, Complete In All Respect Including Wooden Carrying Case. The Standard Microscope Shall Be Equipped With. Inclined Tube At 45 And Rota Table 360 With Perfectly Centered Quadruple Revolving Nose Piece. One 10x Wide Field Eye-Piece. 4x, 10x, 40x SL And 100x SL Oil Immersion.Objectives. Low Drive Coaxial Mechanical With Sub Stage Bright Field Condenser N.A. 1.25 Iris Diaphragm With Built In Swing Out Filter Holder And Up-Down Movement Through Rack Pinion.COAXIEL Coarse Highly Sensitive Fine Movement Mechanical Graduated To 0.02mm. With Locking Device. Heavy Rectangular Sturdy Base With Built In LED COOLED BRIGHT Illumination Or 6V 20 W HALOGEN LAMP With On/Off Switch, Continuously Variable Light Intensity Controlled Regulator Quick Change Bulb Provision And Plano Concave Mirror Attachment Suitable For 220V, 50 Hz, AC.Supplied With Standard Accessories...
Other Specifications Download Specification | |
---|---|
Microscope Type | Pathological Binocular with coaxial focusing control |
Indian Standard Specification | Pathological IS:4381 latest |
ISI MARKED | yes |
C/ML No. | 9474292 |
Type Of Lamp For Illumination | LED/HALOGEN |
Lamp Lumen | 0 lumen |
Light On/Off Switch | yes |
Light Intensity Regulator | yes |
Plano Concave Mirror Attachment | yes |
Limb Type | non Hinged |
Stage | Ractangular |
Size Of Stage | 135mm X150mm |
Coarse And Fine Movement Of Stage | yes |
Eye Piece Type | Compensating |
Eye Piece With Magnification | Set of Two for Binocular 10X |
Binocular Eye Pieces Confirming To The Requirements Of IS: 8275/1976 ( Latest) | Yes |
Co-axial Focussing | yes |
Micrometer Arrangement | No |
Objectives Type | Semiapochromatic |
Numerical Apperture Of Objective | Research 10xNA 0.25 20xNA 0.50 40xNA 0.65 100xNA 1.25 |
Carrying Case | yes |
Warranty | 2 year |
Extra Eyepieces | Set of Two 5X |
Availability Of Test Reports From Central Govt/NABL/ILAC Accredited Lab Covering All The Parameters As Per Specifications | IDDC |
Test Report No. | IDDC:SE(O):468 |
Test Report Date | 10.11.2014 |
Name Of Test Lab | IDDC |
Address Of The Test Lab | STAFF ROAD,AMBALA CANTT |