LCD Display, Pumping Mechanism-Curvilinear peristaltic. Compatible with IV sets of any standard. Flow Rate- 0.1-1500 ml/h in 0.1 ml/h increments. Purge, Bolus 100-1500 ml/h in 1 ml/h increments Purge when pump stops, bolus when pump starts. Bolus volume 1-20 ml in 1 ml increments. Infusion Mode ml/h, drop/min, time-based. KVO Rate 0.1-5 ml/h in 0.1 ml/h increments. Alarms-Occlusion, air-in-line, door open, end program, low battery, end battery,AC power off, motor malfunction, system malfunction, reminder alarm. Additional Features- Real-time infused volume / bolus rate / bolus volume / KVO rate, automatic power switching, mute key, purge, bolus, system memory,key locker, change flow rate without stopping the pump. Occlusion Sensitivity-High, medium, low. Air-in-line Detection- Ultrasonic detector. Wireless management optional. Power Supply, AC-110/230 V optional, 50-60 Hz, 20 VA. Battery-9.6 1.6 V, rechargeable. Battery Life-5 hours at 30 ml/h. Weight- 2.5 kg. Size-174x126x215 mm..
Other Specifications Download Specification | |
---|---|
Material | FIBER |
Technolgy | AUTO |
Capability Of Motioring Of Empty Intravenous Bags Or Bottles | Yes |
Fitting Of Device On The IV Bottle And Bag | Yes |
I.R Sensor To Detect The Abnormilty In I.V Bag Or Bottle | Yes |
Sound Alarm Provided | Yes |
Power Sources | Battery operated |
Details Of Battery In Use | Li ion |
Life Of Battery | 2 years |
Packing | Individual |
Certification | ISO |