Eppendorf Micro Cuvette G1.0 and Eppendorf BioSpectrometer fluorescence - 6137 000.901 (Eppendorf)(Eppendorf Micro Cuvette G1 and Eppendorf BioSpectrometer fluorescence - 6137 000901)

` 696,200.00

  • MRP: `870250.00

A compact UV-Vis spectrophotometer featuring an integrated Fluorescence module. It can be used to increase the measuring range for biomolecules like nucleic acids and proteins by a factor of 1000, compared to UV measurements. Automatic evaluation and storage of measurement data with clear presentation of results. Data transfer via USB interface. Display - VGA TFT Display 5.7 inch, Monochromator - Holographic aberration-corrected concave grating, Wavelength Range user selectable - 200nm to 830nm, Fluorescence excitation 470 nm and emission 520,560 nm, Wavelength Resolution - 0.5 nm, Wavelength Accuracy - /-1nm, Repeatability - 0.5 nm, Spectral Bandwidth - 4nm, Type of Detector - CMOS Photodiode Array, Measuring Modes - Absorbance, concentration, scan absorbance wavelength spectrum, Operating Mode - Pre-programmed and freely programmable, Light Source - Xenon flash lamp. Eppendorf microCuvette G1.0 is a Microvolume measuring cell for measuring 1.5-10 microLiter sample volumes..

  • Availability: 5 In Stock
  • Min. Qty. Per Consignee: 1
  • Display - High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC
  • Monochromator - Concave holographic grating with 1200 lines/mm
  • Wavelength Range (user selectable) - 340-960 nm
  • Wavelength Accuracy - +-1 nm
  • Type of Detector - Photodiode
Other Specifications Download Specification
Spectrophotometer Type Microprocessor Based
Display High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC
Monochromator Concave holographic grating with 1200 lines/mm
Wavelength Range (user Selectable) 340-960 nm
Wavelength Resolution 0.2 nm
Wavelength Accuracy +-1 nm
Repeatability +-1 nm
Spectral Bandwidth < 2
Type Of Detector Photodiode
Measuring Modes Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards.
Operating Mode Single wavelength: Measuring 1, ABS and CONC
Light Source Tungstan Halogen Lamp
Power 230 V +- 10 50Hz
Standard Accessories To Be Supplied 10 nm path length four (4 nos.) Cuvettes matched with 3 percent
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " EN 61010-1 UL 61010-1, CAN/CSA C22.2 No. 61010-1
Power Requirements 230 V +- 10, 50 Hz
Operating Temperature 15 to 35 Degree C
Operating Humidity (RH) 25 to 70 %
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" 2014/30/EU
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) Yes
Test Report No. 20110826-E158089
Test Report Date 26-08-2011
Name Of The Lab Underwriters Laboratories
Warranty 1 Year