Housing Plastic moulded on two wheels Max. Vacuum / LPM -600 mm Hg, 60 Ltrs / min Pump Type Double Diaphragm pump Jars 2 x 2.5 Ltrs. PC Jar with changeover lever Filter Bacterial Filter Autoclavable / Reusable Tubing 10 mm ID x 2 mtr. PVC Vacuum Gauge 6.25 cm dia 0-760 mm Hg Power Supply 220 V AC, 50 / 60 Hz, 180 Watt Noise Level 60 dB A 3 Dimensions 46 x 34 x 86 cms Net Weight 22 Kg Overflow Safety Mechanical / Electronic Acoustic Alarm Accessories / Optional Foot Switch / 110 V on request, 2 Ltr Disposable Jar with Liner Certifications CE, IEC 60601-warranty 2 Year.
Other Specifications Download Specification | |
---|---|
GMDN Name & Code(s) | Suction pump portable & CT 1272 |
"Definition:A Portable Assembly Of Devices Primarily Intended To Be Used By Emergency Medical Services (EMS) To Aspirate Fluids,secretions, Or Other Foreign Materials From A Patient's Airway By Means Of Suction. It Typically Consists Of A Gas Powered Mechanism Driven By Medical Air Or Oxygen (O2) From A Gas Cylinder To Create The Suction (e.g., A Venturi Tube), Tubing, A Collection Container, A Vacuum Gauge And Control Knob, And A Microbial Filter. The Pump Creates A Vacuum In The Suction Tubing Which Is Used For The Removal Of Materials Into The Collection Container. This System Is Typically Used During Patient Transport Or For Emergency Situations." | yes |
Clinical Purpose : To Aspirate Fluids, Secretions, Or Other Foreign Materials From A Patient's Airway By Means Of Suction. | yes |
Technical Characteristics (specific To This Type Of Device) : 0-700 Mm Hg ± 10 Regulable, Flutter Free Vacuum Control Knob, 25ltrs/min, Tight Fitting Jar Cap.Wide Mouthed 2 X 2 Ltrs. (Poly Carbonate) With Self Sealing Bungs And Mechanical Over Flow Safety Device. | yes |
User's Interface | Manual |
Dimensions (Height X Width X Length) Max. | 43 x 30 x 68 cm |
Weight (Max) | 22 Kgs |
Mobility, Portability | yes |
Noise (in DBA) Max. | 60 dBA |
Heat Dissipation: Should Maintain Upto 36.5 Deg Temp And The Heat Disbursed Through A Exhaust Fan | yes |
Operational Requirements : 230 V, 50 Hz, 2 ± 0.5 Amps . | yes |
Protection : Electrical Protection By Resettable Over Current Breakers Or Replaceable Fuses, Fitted In Both Live And Neutral Lines. | yes |
Power Consumption | 200 Watts Watts |
Accessories (mandatory, Standard, Optional);Spare Parts (main Ones) | Foot Switch |
Consumables / Reagents (open, Closed System) | NOT APPLICABLE |
Operating Condition:–Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances. | yes |
Product Should Be FDA / CE Approved Product, ISO 13485:2003; ISO 10079-2-1999: Medical Suction Equipment - Part 1 : | yes |
Warranty | 3 years |
Training Of Staff (Medical, Paramedical, Technician) | yes |