Keysight Technologies DSOX3034T Digital Storage Oscilloscope 350MHz 4CH the InfiniiVision 3000T X-Series redefined oscilloscopes. It saw the most signal detail, provided more functionality than any other oscilloscope, and gave you maximum investment protection. The DSOX3034T Oscilloscope with capacitive touch screen, a user interface designed for touch, and the exclusive hardware base zone touch trigger with two-zone support, all combined with an industry-leading uncompromised update rate of 1 million wfm/s to give you the confidence that you are seeing all of your signal detail and the ability to discover any issues. In addition of new analysis capabilities help you solve your hardest problems quickly. Wide range of serial decodes options, 7-in-1 instrument integration Time/frequency domain correlation and many more. Standard 3 Year warranty. Refer Keysight datasheet for detailed specifications. Standard delivery period 6 weeks. Prices indicated including GST @18 percent...
Other Specifications Download Specification | |
---|---|
Type | Digital |
Operating System | Non -Windows |
Bandwidth Upgradability | Yes |
No. Of Analog Channels | 4 Number |
Digital Channels | No |
No. Of Digital Channels | NA Number |
Record Length Per Digital Channel Simultaneously | NA GS/sec |
Sampling Rate Per Digital Channel Simultaneously | NA Mega Points |
Min. Detectable Pulse Width | NA ns |
Memory Upgradability | No |
Segmented Memory Feature Availability | Yes |
Vertical Sensitivity | 1mV/Div to 5 V/Div |
Vertical Resolution | 8 bits |
Coupling | AC, DC, Ground on Screen |
Time Base Range | 2 ns/div to 50 s/div |
Trigger Types | Mode: Normal, Auto, Single, force Type: Edge, Pulse, video, hardware based zone trigger, runt, setup/hold, pattern, burst etc. |
Measurements | 37 Built-in Automatic Measurement, math function-Two, displays FFT and one math simultaneously. Can be cascaded. Add, subtract, multiply, divide, differentiate, integrate, FFT, Ax + B, squared, square root, absolute value, common logarithm, natural logarithm, exponential, base 10 exponential, low pass filter, high pass filter, averaged value, smoothing, envelope, magnify, max hold, min hold, measurement trend, chart logic bus (Timing or State) |
Frequency Analysis Feature | No |
In-built Frequency Counter | Yes |
In-built Function Generator | No |
Acquisition Modes | Normal, Peak Detect, Averaging, High Resolution, Segmented |
Display Type | Colour |
Display Size | 8.5-Inches Capacitive Touch Display CM CM |
I/O Interface | USB Interface, LAN/ VGA/ GPIB Optional |
Standard Accessories To Be Supplied | Four Nos. of Passive Probe 10:1, Power Chord, Certificate of Calibration, CD Containing Manual |
Operating Temperature Range | 0 to 55 Degree C Degree C Degree C |
Operating Humidity (RH) | Up to 80 RH at or below 40 Degree C % |
Power Supply | 230V +- 10 , 50Hz |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification Including Environmental Tests As Under : 1. Dry Heat Test : 45 Deg C For 16 Hours As Per IS:9000/Pt.3/sec.5/1977 (reaffirmed In 2007) 2. Cold Test : 0 Deg C For 4 Hours As Per IS:9000/Pt.2/sec.4/1977 (reaffirmed In 2007) 3. Damp Heat Test: 95% RH At 40 Deg C, Two Cycles Of 24 Hours Each Separated By A Gap Of 4 Hours As Per IS:9000/Pt.5/sec. 1/1991 (reaffirmed In 2007) The Oscilloscopes Shall Be Checked Functionally Before And After Recovery Period Of 1-2 Hours Of The The Environmental Test Sequence As Above. Functional Tests Shall Include Switching ON And Calibration | yes |
Test Report No. | NTH(SR)EL (C )/ 2013/227G |
Test Report Date | 13/09/2013 |
Name Of The Lab | NTH Taramani, Chennai |
Warranty | 3 Year |
Real Time Bandwidth | 350 MHz |
Record Length Per Analog Channel Simultaneously | Standard 4 Mpts, Standard segmented memory Mega Point |
Sample Rate Per Analog Channel Simultaneously | 5 GSa/s half channels, 2.5 GSa/s all channels GS/sec |
Waveform Update Rate | 1000,000 wfrm/s Waveforms/ |
Time Base Accuracy | 2 PPM |