AAS Spectrophotometer is fully automated and delivers a new level of flexibility and simplicity. It can work as standalone mode for routine analysis and can also be fully PC Controlled. The selected element lamp is automatically aligned in the built-in 6/8 lamp turret which automatically recognizes coded HCL lamps and loads their operating parameters. The instrument provides many features in terms of performance and saftey never before available on low-cost AAS..
Other Specifications Download Specification | |
---|---|
Computer Controlled True Double Beam Atomic Absorption Spectrophotometer | Yes |
Eight Lamp Automatic Turret With Independent Eight Power Supplies With Coded Lamp Compatible | Yes |
Monochromator With Wavelength Range 190-900 Nm. | Yes |
Holographic Grating Of High Density Of Not Less Than 1800 Lines/mm | Yes |
System Have Wavelength Locating By Automatic Peak Searching, Auto Loading Of All Parameters And Auto Bandwidth Selection And Continuously Variable Selection 0.2-2.0 Nm With D2 Lamp Background Correction | Yes |
Titanium Burners For Acetylene Nitrous Oxide (C2H2- N2O) & Acytelene-Air (C2H2-AIR) With Precise Knobs For Burner Optimization I.e. Height, Rotational And Lateral | Yes |
Fully Inert Nebulizer | Yes |
Gas System With Automatic Flame Changeover, Full Safety Interlock Including Pressure Sensors On Both Line, Power Failure Protection, Burner Interlock And Flame Sensor, Flame Ignition Should Be Automatic | Yes |
Burner Movement Thorough Computer, All Three Directional, Height, Rotation Lateral | Yes |
PTFE Spray Chamber And Adjustable Impact Bead Aerosol | Yes |
Single Element Coded Hollow Cathode Lamps : Cu, Fe, Zn, Mn, Mg B, K, Mo. | Yes |
System Have Facilities Of Repeat Of Result Of Same Sample And Date Treatment. Automatic Calculation Of Percentage In Base Material | Yes |
Standard Accessories To Be Supplied : 1. Acetylene And Nitrous Oxide Gas Cylinders Should Have Explosive Certification With Regulator 2. Two Filled Cylinders For Each Gas, 3. Air Compressor, 4. Air Filter, 5. Spares And Consumables For 2 Years Operations. 6. SS Exhaust Fume Hood With Inert Centrifugal Blower, 7. Instruction Manual 8. Circuit Diagram | Yes |
Spare One Set Hollow Cathode Lamps Of Cu, Fe, Zn, Mn. | No |
100 Ml Standard Solutions Of 1000 Ppm For Each Element Under Guarantee Certificate | Yes |
. Training At The Time Of Installation Up To User Level Satisfaction | Yes |
PC Software | Available |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | YES |
Power Requirements | 230 V +- 10, 50 Hz |
Operating Temperature | 15-35 Degree C |
Operating Humidity (RH) | 20-80 % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | YES |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | YES |
Test Report No. | 170902087/170902087-1 |
Test Report Date | 26th September 2017 |
Name Of The Lab | Spectro Analytical Labs Limited |
On-site Warranty | 3 Year |
Availability & Supply Of Spare Parts & Accessories | For next 10 years |