Should be compact weighing 2.5 Kg color TFT display Syringe sizes 5,10,20,30,50/60 ml Should be compatible with syringe of any brand standard Flow rate range Syringe 5 ml 0.1-100 ml/h in 0.01, 0.1, 1, 10 ml/h increments Syringe 10 ml 0.1-300 ml/hSyringe 20 ml 0.1-600 ml/h Syringe 30 ml 0.1-800 ml/h Syringe 50/60 ml 0.1-1200 ml/h Should have Bolus rate5 ml 0.1-100 ml/h in 0.01, 0.1, 1, 10 ml/h increments10 ml 0.1-300 ml/h20 ml 0.1-600 ml/h30 ml 0.1-800 ml/h50/60 ml 0.1-1200 ml/h Anti-Bolus Mechanism Flow rate setting ml/Min, ml/h Should have infusion modes- Flow based, Time based, body weight mode Should have Body weight rate calculation in g/kg/min, g/kg/h, mg/kg/min, mg/kg/hShould have KVO rate 0.1-1 ml/h in 0.01 ml/h increments Should have alarms for Occlusion, Near empty, End programme, low battery, A/C power off, motor malfunction, System malfunction, Pressure sensor error, Syringe drop off..
Other Specifications Download Specification | |
---|---|
Material | FIBRE |
Technolgy | SERVO MECHANISM |
Capability Of Motioring Of Empty Intravenous Bags Or Bottles | Yes |
Fitting Of Device On The IV Bottle And Bag | Yes |
I.R Sensor To Detect The Abnormilty In I.V Bag Or Bottle | Yes |
Sound Alarm Provided | Yes |
Power Sources | Battery operated |
Details Of Battery In Use | LITHIUM ION |
Life Of Battery | TWO YEARS |
Packing | Individual |
Certification | ISO/European-CE |