Spectral Range7800 to 350cm Resolution Better than 0.85 cm Wave number Precision 0.01CM Scanning Speed5- step adjustable for differentAdvantages Large sample compartment.Based on accurate transmission method better than reflection method available with other brand of instruments. Transmission method is more accurate and reliable than reflection method ATR can also be installed OptionalApplicationsSignal to noise ratio better than 15,0001 RMS value, at 2100cm, Resolution 4cm Detector DTGS, 1minute data collection Beam Splitter Ge coated Kbr Infrared Source Air-cooled, hingh efficiency Reflex phere module, Detector DTGS Data System Compatible Computer Software FTIR Software contains all routines needed for basic spectrometer operations, including library search, quantization and spectrum.
Other Specifications Download Specification | |
---|---|
Spectrophotometer Type | Microprocessor Based |
Display | High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC |
Monochromator | Concave holographic grating with 1200 lines/mm |
Wavelength Range (user Selectable) | 340-960 nm |
Wavelength Resolution | 0.2 nm |
Wavelength Accuracy | +-1 nm |
Repeatability | +-1 nm |
Spectral Bandwidth | < 2 |
Type Of Detector | PMT (Photomultiplier Tube) |
Measuring Modes | Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards. |
Operating Mode | Single wavelength: Measuring 1, ABS and CONC |
Light Source | Tungstan Halogen Lamp |
Power | 230 V +- 10 50Hz |
Standard Accessories To Be Supplied | 10 nm path length four (4 nos.) Cuvettes matched with 3 percent |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | ce certificate |
Power Requirements | 230 V +- 10, 50 Hz |
Operating Temperature | 230 Degree C |
Operating Humidity (RH) | 70 % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | NA |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | AS PER OEM |
Test Report No. | AS PER OEM |
Test Report Date | AS PER OEM |
Name Of The Lab | AS PER OEM |
Warranty | 1 Year |