I.V. Cannula with wings without port for infants with a detachable collar on the wings which provides support and facilitates the easy insertion into an infants body. The collar can then be removed thus making the cannula flat for fixing on the body and avoids any interference with the body or accidental removal of cannula when the baby moves.Our I.V. Cannula Range has Needle cover to protect the needle from accidental damage or punctures. Hub Cover i.e., the flashback chamber has high visibility indicating perfect and successful veni-puncture with blood entry. Needles Hub has a holding area precisely designed for placement of thumb. Leur Lock also called stopper used to plug the cannula to prevents any seepage of blood due to back flow, when cannula is not in use. Easy to peel off system to remove form skin blister pack..
Other Specifications Download Specification | |
---|---|
Material Of Needle | Stainless Steel Complying with ISO 9626 |
Catheter | With |
Material Of Catheter | Polytetrafluoroethylene (PTFE) |
Colour Coding And Corresponding Size Of Catheter As Per IS/ISO 10555-5 | Violet - 26G |
Outer Diameter Of Needle As Per IS/ISO 10555-5 | 0.60 mm |
Nominal Outside Diameter Of Catheter Tube In Mm As Per IS/ISO 10555-5 | NA |
Effective Length Of Catheter Tube In Mm As Per IS/ISO 10555-5 | 19 |
Flowrate Of Water Through Catheter In Ml/min As Per IS/ISO 10555-5 | 13 |
Needle Point Finishing As Per IS/ISO 10555-5 | Sharp and Free from Feather Edges, Burrs and Hooks |
Needle Hub | Female Fitting with 6 (Luer) taper Complying with ISO 594-1 |
Injection Port | Without |
Wings | With |
Luer Lock Plug/Cap | Yes |
Radio-detectability Of Catheters | Radio-Opaque |
Vent Fitting | Yes |
Disposable | Yes |
Sterile | Yes |
Non-Toxic And Non-Pyrogenic | Yes |
Blister Packing | Yes |