wave lenght 190-1000 nm.
Other Specifications Download Specification | |
---|---|
Spectrophotometer Type | Microprocessor Based |
Display | High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC |
Monochromator | Concave holographic grating with 1200 lines/mm |
Wavelength Range (user Selectable) | 340-960 nm |
Wavelength Resolution | 0.2 nm |
Wavelength Accuracy | +-1 nm |
Repeatability | +-1 nm |
Spectral Bandwidth | < 2 |
Type Of Detector | Photodiode |
Measuring Modes | Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards. |
Operating Mode | Single wavelength: Measuring 1, ABS and CONC |
Light Source | Tungstan Halogen Lamp |
Power | 230 V +- 10 50Hz |
Standard Accessories To Be Supplied | 10 nm path length four (4 nos.) Cuvettes matched with 3 percent |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | on demand |
Power Requirements | 230 V +- 10, 50 Hz |
Operating Temperature | 90 Degree C |
Operating Humidity (RH) | 90 % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | N/A |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | N/A |
Test Report No. | N/A |
Test Report Date | N/A |
Name Of The Lab | N/A |
Warranty | 1 Year |