Pulse oximeter.
Other Specifications Download Specification | |
---|---|
GMDN Name And Code | Pulse oximeter CT 1446 |
"Pulse Oxymeter, Line Powered:A Mains Electricity (AC-powered) Photoelectric Device Intended For The Continuous Transcutaneous Measurement And Display Of Haemoglobin Oxygen Saturation (SpO2). The Signals, Typically Produced By Light-emitting Diodes (LEDs) And A Receiving Detector In A Probe, Or Directly Built-in, Are Used To Make The Measurements Using The Principle Of Spectrophotometry. The Oximeter Displays The SpO2 Values And May Calculate / Display Other Parameters, E.g., Pulse Rate, Electrocardiogram (ECG). The Device Is Typically Used At The Bedside." | yes |
SpO2 Measurement Range At Least 40-70 And 70 To 99 %, Minimum Gradation 1% | yes |
Accuracy Of SpO2 Better Than ± 1% For Range 40-70 And Better Than ± 3% For Range 70-99. | yes |
Pulse Rate Range At Least 30 To 240 Bpm, Minimum Gradation 1 Bpm | yes |
Accuracy Of Pulse Rate Better Than ± 5 Bpm | yes |
Signal Strength Or Quality To Be Visually Displayed | yes |
Audiovisual Alarms Required: High And Low SpO2 And Pulse Rate (operator Variable Settings), Sensor Disconnected,sensor Failure, Low Battery | yes |
Display Type | TFT Screen |
Memory | Minimum 24 hrs trend memory for SpO2 & PR |
User's Interface | Easily accessible touch button to operate the machine |
Inbuilt Software And/or Standard Of Communication | yes |
Weight | 00 gram |
Mobility, Portability | Protective splashproof case for clean storage and safe transport |
Display Must Allow Easy Viewing In All Ambient Light Levels | yes |
Noise Lavel | 00 dba |
Heat Dissipiation: Should Maitain Nominal Temp And The Heat Should Be Disbursed Through A Exhaust Cooling Fan | yes |
Power Input And Frequency | 220 to 240V, 50 Hz |
Power Consumption | 00 Watt |
Power Cable Length | 3 meter |
Electrical Protection Provided By Fuses In Both Live And Neutral Supply Lines. | yes |
"Operating Condition:Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances. Circumstances" | yes |
"STANDARDS AND SAFETY:Should Be FDA / CE Approved Product ISO 80601-2-61-2011: Medical Electrical Equipment- Part 2-61: Particular Requirements For The Basic Safety Snd Essential Performance Of Pulse Oxymeter. Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1, EMC Safety Confirms To IEC 60601-1-2 Standard Requirement; Manufacturer / Supplier Should Have ISO 13485 Certificate For Quality Standard Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (essential Performance Of Electrocardiographs)" | yes |
Warranty | 3 years warranty & one year with min. 3 free servicing. |
Others: List Of Important Spare Parts And Accessories With Their Part Number And Costing. Certificate Of Callibration And Inspection To Be Provided | yes |