Digital Flame Photometer Dual Channel single aspiration, with filter Na K. Final result in ppm, 2 digit dual LED display for simultaneous result along with compressor accessories, auto ignition, Ca Li filter can be provided on request at extra cost..
Other Specifications Download Specification | |
---|---|
System Is Capable To Measure Alkali And Alkaline Earth Metals (Sodium, Potassium, Calcium, Lithium) In The Samples | Yes |
Ignition System | Auto ignition |
Analyzed Elements | Na+, Ca++, K+, Li+ |
Accuracy | Less than +- 2 % |
Reproducibility | Less than +- 2 % |
Sensitivity | Na+ = 1 ppm, Ca++ = 15 ppm, K+ = 1 ppm, Li+ = 1 ppm |
Display | at least 3 lines LCD |
Detector | Photodiode |
Filters | Sodium, Potassium, Calcium, Lithium |
Aspiration Rate | 3 to 6 ml/min |
Memory | To store and recall 800 samples |
Type Of Printer Interface For Printing Of Samples | Yes |
Accessories | Compressor with built in air regulator and air filter |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | As per OEM |
Power Requirements | 230 V +- 10, 50 Hz AC |
Operating Temperature | As per OEM Degree C |
Operating Humidity (RH) | As per OEM % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | As per OEM |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | As per OEM |
Test Report No. | As per OEM |
Test Report Date | As per OEM |
Name Of The Lab | As per OEM |
Warranty | 1 Year |