CHAMBER SIZE -500 MM DIAMETER X1200 MM DEPTH, 247 LTR. HORIZONTAL CYLINDERICAL HIGH PRESSURE HIGH VACCUM STERILIZER ELECTRICALLY OPERATED WITH DOOR LOCKING FACILITY, LOW WATER PROTECTION SYSTEM, PRESSURE CUT OFF FACILITY AS PER IS3829 PART 2 WITH 6 DIFFERENT SETTING OF THE CYCLE PROGRAMS, DIGITAL DISPLAY CHAMBER, PRESSURE, TERMPERATURE, CYCLE NUMBER, BATCH NUMBER, TIME AND DATE WITH PLC BASED MICROPROCESSOR WITH FACILITY OF HMI WHICH IS INCORPORATED WITH THE STERILIZER, SELECT PLC WITH NUMERICAL TYPE 4 LINE AND 18 KEYS. AUTOMATIC OPERATED THROUGH MICROPROCESSOR PLC, SELECT MAKE, ELECTRIC LOAD - 18 KW PLUS VACUM PUMP LOAD ON 3 PHASE, 4 WIRES, 440 V AC, 50 HZ SUPPLY, WORKING TEMPERATURE 121 TO 134 DEGREE C AND 1.2 TO 2.1 KG/SQUARE CM. FITTED WITH SILICON RUBBER GASKET WHICH CAN WITHSTAND UPTO 140 DEGREE C TEMPERATURE. DOOR, RADIAL LOCKING ARM FOR DOOR, OUTER COVER, INBUILT BOILER, CONNECTING PIPES, JACKET, STAND SS-304 QUALITY.
Other Specifications Download Specification | |
---|---|
Sterilizer, Shallow (Dressing Drum) Conforming To IS: 3831-1979 | yes |
Weather ISI Marked | No |
If Yes, CM/L No. | 0 |
Material Of Body, Lid & Belt And Hasp | Stainless steel |
Designation Of Stainless Steel | 04Ce18Ni10 to IS:6911-1992 |
Finish Of Stainless Steel | 2B as per IS:6911-1972 |
Shape | Cylirical |
Height Of Sterilizer Mm | 350 mm |
Diameter Of Sterilizer Mm | 240 mm |
Avaliability Of Test Report Fron Govt./ NABL / ILAC Laboratory To Prove Conforming To Specification. | NO |
Test Report No. And Date | NO, SIZE OF CYLINDERICAL STERILIZER - 500 X 1200 MM |