Anritsu LMR Master combines a high performance receiver / Spectrum Analyzer with the worlds most advanced handheld Vector Network Analyzer plus a CW / P25 Phase 1 Phase 2, DMR MotoTRBO, TETRA, NXDN, and PTC Signal Generator with internally adjustable power from 0 dBm to -130 dBm. The LMR Master S412E is the ideal instrument for Land Mobile Radio LMR, Professional Mobile Radio PMR technicians and engineers engaged in field testing the RF performance of NBFM, P25, P25 Phase 2 TDMA, DMR MotoTRBO, TETRA, NXDN, dPMR, and LTE for commercial, public safety, maritime, and critical infrastructure radio systems. The base model LMR Master performs testing of narrow band FM radio systems including coverage mapping of RSSI, transmitter SINAD, and transmitter THD. The LMR Master may also be equipped with a Spectrogram Recorder, Interference Analyzer with support for Interference Hunter, Coverage Mapping and GPS Receiver..
Other Specifications Download Specification | |
---|---|
Description Of The Item | Two port vector network analyzers 6GHz |
Frequency Range | 9 kHz to 6 GHz |
Aging Rate Of Spectrum Analyzer Frequency: Better Than 1x10-6 /Year | Yes |
Frequency Counter Resolution(min.) | 1 Hz |
Frequency Span | 0Hz(zero span),1kHz to 6 GHz |
Spectral Purity:SSB Phase Noise @10KHz Offset Shall Be Less Than -90dBc/Hz At 1GHz | Yes |
Spectral Purity :Harmonics @ -40 DBm Level Shall Be Less Than Or Equal To -60dBc | Yes |
Sweep Time @span >or= 1 KHz | 0.1S to 1000 s |
Sweep Time @span = 0 Hz | 10 micro S to 2000 S |
Resolution Bandwidth (-3dB) | 10 Hz to 3 MHz in steps |
Video Bandwidth | 10Hz to 1 MHz in steps |
Amplitude : Max Input Level For Protection | +30 dBm |
Amplitude : Max Input Level For Measurement | +20 dBm |
Amplitude :level Measurement Uncertainity | dB |
Average Noise Level Displayed | 300HzRBW,10Hz VBW: |
Markers | Normal and delta |
Trigger | Free run, video, External |
Display | TFT Colour display with provision for three traces |
Interface | USB and LAN |
Measurement Features | CN ratio,AM,time domain power measurement,channel power measurement,Phase noise measurement,TOI,OBW |
Software For Transfer Of Screen Shot/measurement Data To PC | Yes |
2-port Vector Network Analyzer Functionality ,for General Purpose Testing Applications In The Field,specially For Cable And Antenna Measurements, Is Available. | Yes |
Frequency Range For 2port Vector Network Analyzer | 2 MHz to 6 GHz |
Frequency Resolution (CW ON) For 2port Vector Network Analyzer | 10 Hz |
Return Loss For 2port Vector Network Analyzer | 0.00 to 60 dB |
VSWR For 2port Vector Network Analyzer | 1.01 to 10.00 |
Cable Loss For 2port Vector Network Analyzer | 0.00 to 30.00 dB |
For 2port Vector Network Analyzer: Calibration With Precision Calibration Components-open, Short, Load Should Be Included To Achieve Directivity > 35 DB | Yes |
For 2port Vector Network Analyzer: Distance-to-fault In Co-axial Cables Shall Be Possible | Yes |
Phase Range For 2port Vector Network Analyzer | -180 to +180 degree |
Phase Resolution For 2port Vector Network Analyzer | 0.01 deg |
RF Output Port For 2port Vector Network Analyzer | Type N, female, 50Ohm |
Dynamic Range For 2port Vector Network Analyzer | 60 dB |
Nominal Port Power Output For 2port Vector Network Analyzer:Hi Power/Lo-Power Selectable(power Levels Corresponding To Hi And Lo Levels Shall Be Indicated By Bidder. | High port Power 0, 3 dBm: low port power -25dBm |
Measurement Of S11 & S21 And Smith Chart Measurement Shall Be Possible | Yes |
For 2port Vector Network Analyzer :Vector Voltmeter For Comparision And Verification Of The Phase Of The Phase-matched Cables Shall Be Available | Yes |
For 2port Vector Network Analyzer:Group Delay Measurement Shall Be Possible.Details To Be Declared | Yes |
Operating Temperature | 5 to 40 degree C & 80 RH 40 degree C |
Power Requirement | 230 +- 10 VAC, 50Hz & External DC Input +12.5V to 15V with rechargeable battery of 1500 mAH or higher |
Warranty | 3 years |
OEM Shall Have ISO:14001 Certification For Their Manufacturing Works | yes |
One Set Of Operation And Maintenance Manual Shall Be Supplied | yes |
Standard Accessories Shall Be Supplied | yes |
Availability Of Test Reports From Central Govt/NABL/ILAC Accredited Lab Covering All The Features & Functional Requirements Including Environmental Tests As Per Specifications | yes |
Test Report No | 2160.01 |
Test Report Date | 01.06.2016 |
Name Of Lab | Anritsu Company USA |
Address Of The Lab | 490 Jarvis Drive, Morgan Hill, CA-95037 |