Deep Freezer 300 ltr.
Other Specifications Download Specification | |
---|---|
Conformity To Standards | WHO/PQS/E003/FZ03 |
Size | Small |
Water Pack Freezing Capacity,24hrs | 30 Kg/24 Hrs |
Water Pack Storage Capacity | 292 Kg |
Vaccine Storage Capacity | 300 Ltrs |
Holdover Time | 2.29 Hrs |
Cabinet Type | Chest |
Length Of Cabinet | 663 mm |
Width Of Cabinet | 770 mm |
Height Of Cabinet | 1000 mm |
Cabinet Insultaion | CFC Free |
Colour | Blue |
Ground Clearance | 150 mm |
Material Of Cabinet | SS 304 |
Thickness Of Outer Chamber | .5 mm |
Thickness Of Inner Chamber | .5 mm |
Tube Material | Pure Copper |
Drainage For Waste Water | Yes |
Ambient Temperature | 18 °C |
Internal Cabinet Temperature In Continuous Supply At Ambient Temp +5Deg Cel. To + 45 Deg Cel. | -22 °C |
Temperature Manual Control Facility | Yes |
Refrigerant Cycle Type | Compression Based |
Defrost Mode | Auto |
Refrigerant | CFC Free |
Programmable Micro-processor Control Unit | Yes |
Digital Controller Accuracy | .5 |
Power Supply Voltage | 230 Volts |
Power Supply Frequency | 50 Hz |
Automatic Voltage Stabilizer | Yes |
Power Indiactor | Yes |
Digital Display | LCD |
Audio Alram | Yes |
Visual Alram | Yes |
ON/OFF Switch With Power Indicator | Yes |
Weight | 59 Kg |
Details Of Standard Accessories To Be Provided | 2 baskets , foot |
Spare Parts Availability | Yes |
Compressor Warranty Period | 5 Years |
Machine Warranty Period | 1 Years |
Designed For Tropical Climates | Yes |
FDA Approved | Yes |
CE Approved | Yes |
PQS Status | yes |
PQS Code | who/pqs/e03/fz03.1 |
PQS Validity Date | 5/17/2017 |
Availability Of Test Report From Central Govt./NABL/ILAC Accredited Lab To Prove Conformity To All Declared Specifications | ILAC ACCREDITED LAB |
Test Report No. | WB-14-159 |
Test Report Date | 28/5/14 |
Name Of The Lab | CHINA TESTING & INSPECTION INSTITUTE FOR HOUSEHOLD ELECTRIC APPLIANCES |
Address Of The Lab | NO.3 BOXING BALU , BEIJING ECONOMIC AND TECHNOLOGICAL DEVELOPMENT ARE BEIJING 100176?P.R. CHINA NO 29 XIA XIE STREET, XICHENG DISTRICT? BEIJING 100053?P.R. CHINA |
Gross Volume (for Small Capacity It Must Be In Range 105-125Ltrs , And For Large Capacity It Must Be In Range Of 275-300 Ltrs As Specified By WHO/PQS Standards.) | 300 Ltrs |