INTRODUCTIONNATIONAL HORIZONTAL CYLENDRICAL AUTOCLAVE is based on the principle that microbiological actionof saturated steam at elevated temperature is rapid and through All models work on the principle ofdownward displacement of air which is the most economical method of obtaining sterilization Ro bust andrigid construction, designed for all types of bulk sterilization which are commonly needed in Medical,Agricultural, research Institute and Pharmaceutical Industries Suitable the for Sterilizing hospital dressings,rubber, plastic goods, surgical instruments, glass wares, etcCONSTRUCTIONTriple walled with steam jacket and separate boiler Inner chamber and steam jacket are made up of Heavygauge S S 316/30 4 Sheet with leak proof argon-arc welding The sterilizer has single piece door made ofstainless ste el 316/304 Back plate and ring is also made of thick stainless a steel sheet All the sterilizers arehydraulically tested to withstand 25kg/cm2 times the working pressure M ounted on tubular steel framewith ground leveling screwed langes The Outer jacket is wrapped with asbestos sheet or glass wool tominimize the heat losses due to radiation and is covered by polished stainless steel sheet for that elegantappearan ceSTEAM GENERATORMade of heavy stainless steel sheet Heavy ring mounted in front of the boiler with folding thick stainlessst eel plate is itted with heating elements and low water level cutoff device to protect the former fromburning out dry Front folding plate system provides for easy cleaning of the deposited scale on theelements and the walls, for long life and eficiency Fitted with water gauge glass for water level indication,water inlet and outlet valves PRESSURE CONTROLThe PIEZOSTAT Pressure control device is incorporated in all electrically operated sterilizers Iteconomizes on power consumption and reduces the frequent opening of the steam release valve and thusprevents the release of steam in the room, It cuts off the power supply to heaters when the set pressure isachieved and re-energies the heating elements when pressure falls below the set pointHEATING ELEMENTSFlanged type, immersion heating elements is made of high grade material and duly ISI markedSAFETY DOORSterilizer has hinged self-locking single piece thick SS 316/304 plate door and cannot be opened whenunder pressure The self-locking device automatically disengages the thread mechanism when thesterilizer is under pressure to preclude any eventuality of accidentSINGLE POINT CONTROLComplete sterilizing cycle is controlled from one point with the help of Multiport Valve itted at front topTwo speeds of steam exhaust are available fast and slowEasily readable jacket and chamber pressure on gauges mounted on the multi port valve A moisture trapis itted in the chamber discharge line to absorb the condensate automatically to prevent moistening of thesubject matter to be sterilized CONTROL PANEL SWITCH BOXTo provide electrical control panel box and mounted on the stand of the sterilizer for easy operationElectri cal parts pressure switch, contactor, water level relay, MCB, YBR Indicators, on off switchOTHER FEATURES TRIPLE SAFETY The Sterilizer is provided with the triple safety features At the boiler level by aPIEZOSTAT which automatically limits the pressure to the required set value and a spring loaded safetyvalve in case of its failure which releases the steam to keep pressure within the safety limits At thechamber level, a spring loaded safety valve and a dead weight release valve to release the steam in case thepressure exceeds the safety limit All these safety features function independent of one another andsequentially ie one tak e over the charge in case of others failure A plug screen prevents the discharge line from choking The plug i s easily removable for daily cleaning Powerful Ejector for drying sterilized linen circulates air throughout the chamber The circulating airpasses through a corrosion resistant metallic wool ilter A dial thermometer is provided to show the chamber temperature Automatic Vacuum breaker is provided to break vacuum in case of formation of vacuum due to steamcondensatio nOPERATING TEMP PRESSURESterilizing Temp 121 C t o 134CSterilizing pressure 12 to 2 kg/cm2 15 psi to 22psiPower Requirement Suitable to operate on 440V volts, 3 ph, 50 Hz, Ac supply.
Other Specifications Download Specification | |
---|---|
Sterilizer, Shallow (Dressing Drum) Conforming To IS: 3831-1979 | yes |
Weather ISI Marked | No |
If Yes, CM/L No. | N.A |
Material Of Body, Lid & Belt And Hasp | Stainless steel |
Designation Of Stainless Steel | 04Ce18Ni10 to IS:6911-1992 |
Finish Of Stainless Steel | 2B as per IS:6911-1972 |
Shape | Cylirical |
Height Of Sterilizer Mm | 350 mm |
Diameter Of Sterilizer Mm | 225 mm |
Avaliability Of Test Report Fron Govt./ NABL / ILAC Laboratory To Prove Conforming To Specification. | NABL |
Test Report No. And Date | NABL/024/2016 |