Throughput of 240 tests / hour photometric tests single/double reagent Refrigerated reagent/sample compartment On-board laundry 12 steps rinsing system Reusable cuvettes Rear spectrophotometry of holographic concave flat field grating Syringes are made of long life high-precision ceramic piston Collision protection, self-resetting probe Daily and monthly QC LIS interface User authority hierarchical management to enhance system security Sample Reagent Unit - Sample reagent positions- The reagent and sample share one disk, totally 67 positions. Continuously cooling at all positions to keep 5150C within 24 hrs. Sample volume- 3 - 35 L with increments of 0.1 L Sampling technology- Liquid level detection and collision protection Automatic sample dilution-2 115 times. Grating Photometry Grating Photometry Net Weight 120kg.
Other Specifications Download Specification | |
---|---|
System Is Capable To Measure Alkali And Alkaline Earth Metals (Sodium, Potassium, Calcium, Lithium) In The Samples | Yes |
Ignition System | Auto ignition |
Analyzed Elements | Na+, Ca++, K+, Li+ |
Accuracy | Less than +- 2 % |
Reproducibility | Less than +- 2 % |
Sensitivity | Na+ = 1 ppm, Ca++ = 15 ppm, K+ = 1 ppm, Li+ = 1 ppm |
Display | at least 3 lines LCD |
Detector | Photo conductive cell |
Filters | Sodium, Potassium, Calcium, Lithium |
Aspiration Rate | 3 to 6 ml/min |
Memory | To store and recall 800 samples |
Type Of Printer Interface For Printing Of Samples | TCP/IP network connection, standard RS-232C |
Accessories | Compressor with built in air regulator and air filter |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | YES |
Power Requirements | 230 V +- 10, 50 Hz AC |
Operating Temperature | 2-30 Degree C |
Operating Humidity (RH) | 80% relative humidity % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | N/A |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | YES |
Test Report No. | yes |
Test Report Date | YES |
Name Of The Lab | YES |
Warranty | 1 Year |