AC/DC Universal SuctionCategories Medcial Electric Equipment 7A-23 D AC/DC Universal SuctionAdopting oil-free lubrication pump, square negative pressure dial and built-in bottle. Low noise, small size, easy to carry. It adopts three modes to supply electricity alternating current, direct current and batteries. So it is more suitable for lack of alternating current situation such as on vehicles and outdoor places to save and nurse patients.Power supply AC 220V1050Hz212VDCPower90VAMax negative pressure value0.075MPaPumping rate15L/minReservoir volume 2500mLNoise65dB AWe are highly engaged in manufacturing and supplying high quality of Portable Suction Machine.FeaturesSupreme strengthEasy installationEfficient working.
Other Specifications Download Specification | |
---|---|
GMDN Name & Code(s) | Suction pump portable & CT 1272 |
"Definition:A Portable Assembly Of Devices Primarily Intended To Be Used By Emergency Medical Services (EMS) To Aspirate Fluids,secretions, Or Other Foreign Materials From A Patient's Airway By Means Of Suction. It Typically Consists Of A Gas Powered Mechanism Driven By Medical Air Or Oxygen (O2) From A Gas Cylinder To Create The Suction (e.g., A Venturi Tube), Tubing, A Collection Container, A Vacuum Gauge And Control Knob, And A Microbial Filter. The Pump Creates A Vacuum In The Suction Tubing Which Is Used For The Removal Of Materials Into The Collection Container. This System Is Typically Used During Patient Transport Or For Emergency Situations." | yes |
Clinical Purpose : To Aspirate Fluids, Secretions, Or Other Foreign Materials From A Patient's Airway By Means Of Suction. | yes |
Technical Characteristics (specific To This Type Of Device) : 0-700 Mm Hg ± 10 Regulable, Flutter Free Vacuum Control Knob, 25ltrs/min, Tight Fitting Jar Cap.Wide Mouthed 2 X 2 Ltrs. (Poly Carbonate) With Self Sealing Bungs And Mechanical Over Flow Safety Device. | yes |
User's Interface | Manual |
Dimensions (Height X Width X Length) Max. | 43 x 30 x 68 cm |
Weight (Max) | KG Kgs |
Mobility, Portability | yes |
Noise (in DBA) Max. | dba dBA |
Heat Dissipation: Should Maintain Upto 36.5 Deg Temp And The Heat Disbursed Through A Exhaust Fan | yes |
Operational Requirements : 230 V, 50 Hz, 2 ± 0.5 Amps . | yes |
Protection : Electrical Protection By Resettable Over Current Breakers Or Replaceable Fuses, Fitted In Both Live And Neutral Lines. | yes |
Power Consumption | 200 Watts Watts |
Accessories (mandatory, Standard, Optional);Spare Parts (main Ones) | - |
Consumables / Reagents (open, Closed System) | - |
Operating Condition:–Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances. | yes |
Product Should Be FDA / CE Approved Product, ISO 13485:2003; ISO 10079-2-1999: Medical Suction Equipment - Part 1 : | yes |
Warranty | 3 years |
Training Of Staff (Medical, Paramedical, Technician) | yes |