Range- 190-1100 nm, supplied with software for interfacingWavelength range190 - 1100 nmSelectable wavelength0.1 nm incrementsscan speed3000 to 2 nm/minLight source changeAuto change from 300 to 360 nm wavelength range, light source can be set at 0.1 nmadjustments to auto shift light source from UV to Visible lamp or vice-versa.Measurement methodDouble beamLight source Halogen 20W Deuterium LampPhotometric systemCzerny-Turner mountingDetectorSilicon photodiodeSpectral bandwidth1 nm 190 to 1100 nmWavelength accuracy 0.3 nm 0.1 nm at 656.1 nmWavelength repeatability0.1 nmStray lightLess than 0.02Photometric rangeAbsorbance -4 to 4 AbsTransmittance 0.0 to 400Photometric accuracy0.004 Abs at 1.0 Abs0.002 Abs at 0.5 AbsPhotometric repeatabilityLess than 0.001 Abs at 1.0 AbsDriftLess than 0.0003 Abs/hBaseline flatnessLess than 0.0006 AbsNoiseLess than 0.00005 AbsDimensions W x D x H450 x 500 x 275 mm or lessWeight15 Kg.
Other Specifications Download Specification | |
---|---|
Spectrophotometer Type | Microprocessor Based |
Display | High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC |
Monochromator | Concave holographic grating with 1200 lines/mm |
Wavelength Range (user Selectable) | 340-960 nm |
Wavelength Resolution | 0.2 nm |
Wavelength Accuracy | +-1 nm |
Repeatability | +-1 nm |
Spectral Bandwidth | |
Type Of Detector | Photodiode |
Measuring Modes | Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards. |
Operating Mode | Single wavelength: Measuring 1, ABS and CONC |
Light Source | Tungstan Halogen Lamp |
Power | 230 V +- 10 50Hz |
Standard Accessories To Be Supplied | 10 nm path length four (4 nos.) Cuvettes matched with 3 percent |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | yes CE certified |
Power Requirements | 230 V +- 10, 50 Hz |
Operating Temperature | 15-35 Degree C |
Operating Humidity (RH) | 30 to 80% % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | 03 |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | ce certificte |
Test Report No. | ZEAE-1058A(1/3) |
Test Report Date | 30/09/2008 |
Name Of The Lab | CE LAB |
Warranty | 1 Year |