× All the transactions performed on 2.0 is available here. To access Market click "Home".

thermostatic water bath (pacific)(pac-twb)

` 12,330.00

  • MRP: `13700.00

Double walled, inside of heavy gauge high grade Stainless Steel. Outside of Mild Steel. High quality Kanthal make, easily replaceable heating elements are placed in the chamber. Temperature range from 5 C above ambient temp. to 100 C. Temperature uniformity -0.1C . Controlled by digital temperature controller Suitable for 6 Nos. flasks. Adjustable rings for flask positioning with Lids...

  • Availability: 6 In Stock
  • Min. Qty. Per Consignee: 2
  • Minimum numbers of thermometers which can be simultaniously calibrated in water bath. - 20 Nos
  • Reference thermometer confirming to IS:4825/1982 - YES
  • Min. resolution of reference thermometer - 0.01 °C
  • The Comparator water bath either screw type or bubble type as per IS:6274/1971,should be constant and uniform within ±0.01 °C,stability ±0.01 °C and should have digital indication with resolution of 0.01 °C. Thermocouple shall be calibrated and shall have calibration certificate. - Yes
  • Centrifuge to drive mercury level well below 25C at 600 rpm. - Yes
Other Specifications Download Specification
Purpose Of Waterbath. To test clinical thermometer (Liquid in glass(LIG) &electrical) with maximum device within accuracy of ± 0.15 C
Minimum Numbers Of Thermometers Which Can Be Simultaniously Calibrated In Water Bath. 20 Nos
Reference Thermometer Confirming To IS:4825/1982 YES
Min. Resolution Of Reference Thermometer 0.01 °C
Min.uncertainty Of Reference Thermometer(at Coverage Factor K=2) 0.02 °C
The Comparator Water Bath Either Screw Type Or Bubble Type As Per IS:6274/1971,should Be Constant And Uniform Within ±0.01 °C,stability ±0.01 °C And Should Have Digital Indication With Resolution Of 0.01 °C. Thermocouple Shall Be Calibrated And Shall Have Calibration Certificate. Yes
Mechanism To View The LIG Thermometer With Magnification Of 10X. Yes
Mechanism To Hold Thermometer In Water For At Least 20 Seconds. Yes
Centrifuge To Drive Mercury Level Well Below 25C At 600 Rpm. Yes
Thermometer Calibrator Shall Be Capable To Calibrate The LIG Thermometer As Per IS:12622/1989 Yes
All The Gazette Of India Related To Legal Metrology Issued Time To Time Are Applicable To The Specification As Well As To The Product.Also As Per The Gazette Of India Extraodinary,The Legal Metrology Act 2009,cl.22, Approval Of Model Before Manufacturing And Cl.23.Prohibition On Manufacture,repair Or Sale Without Licence For Product Shall Be Applicable On Manufacture Of Measuring Instrument. Yes
Availability Of Calibration Test Reports Of The Reference Thermometer From NPL Or Central Govt/NABL/ILAC Accredited Lab . Yes
Availability Of Type Test Reports Of The Product From Central Govt/NABL/ILAC Accredited Lab Showing Compliance To The Specification Yes
Test Reports No. AS PER OME
Test Reports Date AS PER OME
Name Of The Labs AS PER OME
Address Of The Lab AS PER OME
Product Code Available On Supplier Website Yes
Product Code No PAC-TWB

Price Dekho Price Dekho Price List in india