Inner Chamber is made up of thick non-magnetic steel of 304/316 quality and jacket is made out of rolled steel.Single door made of rolled steel lined with S.S. sheets has radial locking system.All units are hydraulically tested upto 40psi.The space between outer cover made of thick rolled steel sheet and the steam jacket is filled with asbestos sheet or glass wool to minimize thermal losses.Boiler is made of thick stainless steel sheet and is located under the sterilizer.Unit is fitted with water level indicator, safety valve, pressure gauge, water inlet and drain valves.Unit is also fitted with low-water level cut-off device and automatic pressure control device.All MAC rectangular sterilizes are fitted with SELF LOCKING SAFETY DOOR device.The door cannot be opened when chamber is under pressure. Two separate valves are provided for evacuating steam from inner chamber and jacket, as and when desired.Fitted with safety valve as a safety device..
Other Specifications Download Specification | |
---|---|
Sterilizer, Shallow (Dressing Drum) Conforming To IS: 3831-1979 | yes |
Weather ISI Marked | yes |
If Yes, CM/L No. | N/A |
Material Of Body, Lid & Belt And Hasp | Stainless steel |
Designation Of Stainless Steel | 04Ce18Ni10 to IS:6911-1992 |
Finish Of Stainless Steel | 2B as per IS:6911-1972 |
Shape | Cylirical |
Height Of Sterilizer Mm | 350 mm |
Diameter Of Sterilizer Mm | 240 mm |
Avaliability Of Test Report Fron Govt./ NABL / ILAC Laboratory To Prove Conforming To Specification. | ERTL(N)/90(4)-2KA0368 |
Test Report No. And Date | 15.9.2017 |