× All the transactions performed on 2.0 is available here. To access Market click "Home".

single beam uv vis spectrophotometer (labtronics)(lt-291)

` 166,200.00

  • MRP: `188950.00

Range 195 to 1000 nm with facility for automatic concentration, Transmission, Absorbance K Factor measurement. Automatic wavelength calibration and selection. It has sample holder with 4 position selector control for 10 mm path length cuvettes. Automatic wavelength calibration wavelength scan. Automatic peak picking spectrum display. Spectrum and data can be printed out by printer and sent to computer via USB port. Set of 2 Quartz 4 Glass Highly stable and accurate. SOFTWARE INCLUDED.

  • Availability: 9 In Stock
  • Min. Qty. Per Consignee: 1
  • Display - High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC
  • Monochromator - Concave holographic grating with 1200 lines/mm
  • Wavelength Range (user selectable) - 340-960 nm
  • Wavelength Accuracy - +-1 nm
  • Type of Detector - Photodiode
Other Specifications Download Specification
Spectrophotometer Type Microprocessor Based
Display High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC
Monochromator Concave holographic grating with 1200 lines/mm
Wavelength Range (user Selectable) 340-960 nm
Wavelength Resolution 0.2 nm
Wavelength Accuracy +-1 nm
Repeatability +-1 nm
Spectral Bandwidth < 2
Type Of Detector Photodiode
Measuring Modes Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards.
Operating Mode Single wavelength: Measuring 1, ABS and CONC
Light Source Tungstan Halogen Lamp
Power 230 V +- 10 50Hz
Standard Accessories To Be Supplied 10 nm path length four (4 nos.) Cuvettes matched with 3 percent
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " CE2597
Power Requirements 230 V +- 10, 50 Hz
Operating Temperature 00 Degree C
Operating Humidity (RH) 00 %
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" 00
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) 00
Test Report No. 00
Test Report Date 00
Name Of The Lab 00
Warranty 1 Year

Price Dekho Price Dekho Price List in india