800x480TFT Color LCD shows the working status and ECG waveform Touch Screen and soft keyboard control, more convenient for operation.12 Leads ECG simultaneous acquisition, digital signal processor, to get Higher quality ECG waveforms through the AC filter, baseline filter and EMG Filter of the ECG signals.Simultaneous display of 3/6/12 leads ECG waveform, and the state of Printing mode, sensitivity, speed, filter etc, to be easy for interpretation.Sampling twelve leads ECG signal at the same time, analyzing 2 61rhythm lead, 2 6, 3 4, 3 41rhythm lead, 4 3, 4 31rhythm Lead, 6 2, 6 21rhythm leadrecording mode at the same time, and Reporting with multi-formats.The power supply includes both AC/DC. This device includes built-in lithium Polymer rechargeable battery. It can standby for 4 hours, print 150 pieces Of ECG waveform and print 90 minutes continually under the best DC state To satisfy the need of visiting patients and physical examination.It has build-in sto.
Other Specifications Download Specification | |
---|---|
GMDN Name And Code | Multichannel Electrocardiographic , CT-1115 |
Functional Requirements: Electricity (AC-powered) Bedside Device Designed To Continuously Detect, Measure, And Display A Patient's Electrocardiogram (ECG) Through Leads And Sensors Attached To The Patient, It Also Typically Display Heart Rate. The Device Is Typically Equipped With Audible And/or Visual Alarms That Are Triggered When The Patient's Parameters Drop Below Or Exceed Pre-set Limits. Continuous Display Of Patient ECG And Heart Rate On Screen,Allows Display Of Single, 5 Lead ECG Or Simultaneous Display Of At Least 5 Waves Selected From Up To 12 Points.Operator Can Set Audiovisual Alarm Levels For Low Or High Heart Rate.Operates From Mains Voltage Or From Internal Rechargeable Battery. | yes |
Heart Rate Measurement Range To Be At Least 30 To 250 Bpm, With Accuracy Better Than ± 5 Bpm | yes |
Heart Rate Trend Display Of At Least Previous 24 Hours | yes |
Arrhythmia Detection Facility Required; Minimum Gradation Of 1 Bpm | yes |
Audiovisual Alarms Required: High And Low Heart Rate (operator Variable Settings), Cardiac Arrhythmia, Sensor / Wire Disconnected, Low Battery | yes |
User's Interface | manual |
In Built Software And/or Standard Of Communication | yes |
Weight | 1400 gram |
Mobility, Portability | Protective splashproof case for clean storage and safe transport |
Display Must Allow Easy Viewing In All Ambient Light Levels | yes |
Noise Lavel | 50 dba |
Heat Dissipiation: Should Maitain Nominal Temp And The Heat Should Be Disbursed Through A Exhaust Cooling Fan | yes |
Power Input And Frequency | 220 to 240V, 50 Hz |
Power Consumption | N/A watt |
Power Cable Length | 3 meter meter |
Electrical Protection Provided By Fuses In Both Live And Neutral Supply Lines. | yes |
"Operating Condition:Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances" | yes |
"STANDARDS AND SAFETY:FDA / CE Approved Product; Manufacturer / Supplier Should Have ISO 13485 Certificate For Quality Standard. Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (essential Performance Of Electrocardiographs)" | yes |
Warranty | 3 years warranty & one year with min. 3 free servicing. years |
Others: List Of Important Spare Parts And Accessories With Their Part Number And Costing. Certificate Of Callibration And Inspection To Be Provided | yes |