Thermographic IR camera Sonel KT-400 with high-quality, rugged and ergonomic design, brief operation interface and intelligent button navigation KT-400 is equipped with optional lens set for different FOV requirements in different occasions. With built-in illuminator, you can easily take visible photos and compare them with IR photos, which help you find problems more easily. 4 high light touch screen with adjustable brightness. Still clear under sunshine. All operation can be completed on the touch screen. Optional Lens set, you can choose the proper lens for different occasions. Built-in illuminator- Take visible photos in the low lighting environment Different image typeIR, visible Removable SD card- 32GB storage capacity for photos and videos compression storage Wi-Fi share Share IR images and findings immediately Connected with mobile terminal. Connected with mobile terminal to take IR photo, to achieve multi screen control..
Other Specifications Download Specification | |
---|---|
Type | Hand held |
Detector Resolution (min.) | 1,00,000 Pixels |
Temperature Range Of Objects (Degree Celsius) | -20 to +500 °C |
Temperature Measuring Accuracy Of Reading (or Better) | ± 2.0 % °C |
Minimum Focus Distance | N ≥ 10 cm |
Colour LCD Display | N ≥ 8.75 cm |
Image Capture Frequency/ Frame Rate | N ≥ 9 Hz |
Image Detector Type | Uncooled focal plane array (FPA)Vox/ Asi/BST/Silicon |
Thermal Sensitivity (NETD) | N ≤ 0.08 °C |
Submit OEM Certificate Or Certificate From Detector Manufacturer For Thermal Sensitivity (NETD ) | Yes |
Infrared Spectral Band (or Better Coverage Range.) | 8≤ N≤ 14 um |
Detector Resolution | N ≥ 100000 Pixels |
Image Palettes | N ≥ 4 |
Image Capture, Review And Save Mechanism | One handed image capture, review and save capability |
Storage Medium:- Mini SD Card/SD Memory Card/ USB Stick Of 2 GB (min.) | SD Memory Card and USB stick of 16GB and support up to 32 GB |
File Formats ( JPEG Format Or Enhanced Format ) | JPEG/JPG with measured temperature data info, image can be displayed on PC with temperature data info. |
Built-in Digital Camera/ Visual Camera Resolution | N ≥ 1 Mpixels |
Horizental Field Of View Of Standard Lens | N ≥ 20 Degree |
Vertical Field Of View Of Standard Lens | N ≥ 14 Degree |
Horizental Field Of View Of Telephoto Lens | N ≥ 9 Degree |
Vertical Field Of View Of Telephoto Lens | N ≥ 7 Degree |
IFOV Of Standard Lens | N ≤ 2.1 mRad |
IFOV Of Telephoto Lens | N ≤ 1.0 mRad |
IFOV Of (Spatial Resolution) | N ≤ 1.4 mRad |
User Selectable Emissivity Setting/correction | 0.01 ≤ N≤ 1 |
Details Of Operating Functions | Wi-fi Image Transfer and sharing, multi screen control, Center point temperature, Auto tracking of max. & min. temperature, Alarm max. temperature, Alarm min. temperature, Alarm function for min. & max. temperature and low battery indication, dew point temperature, instant report generation |
Operating Temperature ( 0 To 50 Degree Celsius) | -10°C to 50°C °C |
Control And Adjustments:- The Set Up Menu Should Provide The Facility For The Adjustment/ Selection Of The Time/date Etc. | Yes |
Software:- With Compatible Software And Hardware For Downloading On PC With Windows(Latest) | Yes |
Image Output:- Image Shall Be Able To Be Viewed On PC Or Readable Through Mini SD Card/SD Memory Card/USB Stick Storage Medium | Yes |
Battery Back Up 3hrs Min With Built-in Suitable Charger Or Charger Separately Provided | Li-Ion rechargeable battery runs up to 4 hours operation and inbulit camera charger hrs |
Details Of AC Battery Charging | Power adapter 110/230 VAC, 50/60 Hz facilitate in camera charging |
Power Management:- Automatic Shut Down /User Selectable | Auto Power off Sleep mode/User selectable |
UL Or CE Certificate For Safety | Yes |
Size/ Dimension | 264 mm x 137 mm x 78 mm |
Weight (including Batteries) | 740gm |
IP Protection (IP 54 Or Better) | IP53 |
Focus | Manual |
Interface | micro USB, HDMI, SD Card |
Cost Of Calibration Beyond Warranty | 10 % of purchase cost |
Details Of Voice Annotation (if Any, Not Mandatory) | 60 sec voice record |
Thermal Imagers Are Meant For Preventive Maintenance Purpose Such As Detecting Heat Build Up On Bus Bars, Bus Ducts And Isolators Etc, Heat Generation Detection Of Sub Stations And Control Panels Of HT And LT Lines, Scanning Of Bushing Of Transformers, Scanning Of Circuit Breakers, Scanning Of Joints In 25KV/33KV Overhead Equipments. Scanning Of All Types Of Rotating Machines, Firefighting Help, Building Diagnostics Etc. | Yes |
Submit Calculation, Showing Minimum Size Of Object, Which Thermal Profile Can Be Scanned By Imager Within Specified Limit Of Accuracy From Maximum Distance To Have First Hand Information With Respect To Its Utility From User's Point Of View. | Yes |
Software Along With The Equipment | Inclusive |
Adequate Facility For Verification Of Features (excluding IP Protection)of System. | Yes |
Availability Of The Test Report From Central Govt./ NABL/ ILAC Acredited Lab As Per Doc Correction Slip No. 38 Dtd 18/12/12 | Yes |
Test Report No., Date Of Test Report Covering Verification Of All Features & Functional Parameters & Environmental Tests Sequence For "one Sample Of Each Type/range" Of Thermal Imager. | ST/CAL-01 08-Jan-2018 |
Name Of Lab | Calibration/test certificates are provided |
Address Of Lab | Calibration/test certificates are provided |
Type Test Reports Shall Consist Of Verification Of All The Features & Functional Parameters & Environmental Tests Sequences; A. Dry Heat Test (For 16 Hrs At A Temp. Of 55 Degree C In Accordance With IS: 9000/part-3/section-5/1977) (reaffirmed In 2004). B. Cold Test(For 4 Hrs At A Temp. Of -20 Degree C In Accordance With IS: 9000/part-2/section-4/1977 (reaffirmed In 2004). C. Damp Heat Cyclic Test ( For 2 Cycles Of 24 Hrs At A Temp. 40 Degree C & 95% RH In Accordance With IS: 9000/part-5/section-1/1991 (reaffirmed In 2004). | Yes |
After Environmental Test Sequence As Above And A Recovery Period Of Two Hours,the Equipment Shall Be Checked For Complete Parameters, Excluding IP Protection. | Yes |
Type Test Reports Shall Consist Of Verification Of All The Features & Functional Parameters &Operating Temperature Tests Sequences; A. The Equipment Shall Be Kept At Zero Deg Celsius For At Least 2 Hours And Measurement Of Object Temperature Shall Be Done To Cover The Entire Temperature Range Specified. B. The Equipment Shall Be Kept At 50 Deg Celsius For At Least 4 Hours And Measurement Of Temperature Shall Be Done To Cover The Entire Temperature Range Specified. C. Drop Test Of Unpacked Thermal Imager From A Height Of 2m. D. IP Protection Test. (Test Method Shall Be As Per Relevant IS: 9000 Series Of Specification) | Yes |
Availability Of Master Calibrator And Source To Verify The Calibration Of The Instrument On 10% Samples. | Yes |
Facility Of Environmental Chamber Of Temperature Range 0 To 50 Deg Celsius Or Better To Assess The Operation Of The Equipment In Operating Condition | Yes |
Availability Of Facility For Verificaion Of All Features & Functional Performance Of The Equipment Except IP Protection On 10% Samples. | Yes |
Furnished Optical Calculation To Have Idea About Distance To Spot Size Ratio. | Yes |
Instructions To Maintain Safe Distance From Heated / Live Objects While Capturing The Image. | Yes |
The Infrared Imaging System Shall Be A Single Unit Suitable For Hand Held Operation | Yes |
The System Shall Be Capable Of Working In Open Environment And Shall Not Use Any Cooling System Either Internally Or Externally | Yes |
The Accessories Shall Include One Sets Of Spare Rechargeable Batteries, AC Power Adapter, Battery Charger, Carrying Case, User Manual And Software Installation CD And Any Other Application S/W That May Be Essential To Make The System Complete In All Respect. | Yes |
In Typical Ambient Environments, When The Cameras On/off Switch Is Activated To Apply Power To The Camera, The Camera Shall Render A Usable Thermal Image In One Minute Or Less. | Yes |
Camera Shall Not Incorporate A Sleep Mode That Turns-off The Cameras Internal LCD Display While The Camera Is "ON". | Yes |
The Camera Shall Incorporate An On-screen Battery Level Indicator | Yes |
The Camera Shall Incorporate An Automatic Active Temperature Stabilization System By Use Of Suitable Method. When The Cameras Electronics Heat-up To A Predetermined Critical Temperature, The Temperature Stabilization System Automatically Activates To Dissipate Heat From The Electronics And Move The Heat Into A Heat Sink Or Camera Shall Be With Uncooled Microbolometer (OEM Certificate Shall Be Furnished To This Effect) | Yes |
The Camera Shall Incorporate Through-the-lens Direct Temperature Measurement Capability That Measures Relative Surface Temperature Readings. The Direct Temperature Measurement Readings Shall Be Displayed, In Numerical Form. | Yes |
The Camera Shall Be Fitted With Li Ion Or Better Chemistry Type Rechargeable Batteries | Yes |