Syringe PumpIt can set Bolus rate from 1ml/hr to 1200ml/hr in increments of 1ml/hr. should have facility to set, bolus volume in the increment of 1ml.Keep Vein open KVO available with a facility to set KVO flow rates and option to keep the function OFF. Range 0.1 to 3ml/hrShould detect free flow or line disconnection.Should have facility to store the last infusion data.Pressure monitoring system for occlusion alarm. Selectable occlusion pressure trigger levels from 100-900 mmHg with a choice to select the default setting by the operator in steps of 50 mmHg is must. Facility to display the actual pumping pressure in numeric / graphical form is required.Have comprehensive alarm package including pre occlusion occlusion limit exceed alarm, Near end of infusion pre-alarm and alarm, Volume limit pre -alarm alarm, KVO rate flow, Low Battery pre alarm and alarm, AC power failure, Drive disengaged, Infusion Line disconnection alarm etc.Works on standard disposab.
Other Specifications Download Specification | |
---|---|
Material | Electric Plastic & Metal |
Technolgy | Micro controller system |
Capability Of Motioring Of Empty Intravenous Bags Or Bottles | Yes |
Fitting Of Device On The IV Bottle And Bag | Yes |
I.R Sensor To Detect The Abnormilty In I.V Bag Or Bottle | Yes |
Sound Alarm Provided | Yes |
Power Sources | Battery operated |
Details Of Battery In Use | Sealed NiMH Battery having 10hr backup for about 5ml/hr flow rate with 50ml syringes |
Life Of Battery | Sealed NiMH Battery having 10hr backup for about 5ml/hr flow rate with 50ml syringes |
Packing | Individual |
Certification | ISO / CE |