needle destroyer manualy with out electronic.
Other Specifications Download Specification | |
---|---|
Purpose | To destroy the used injection needles by instant electromelting process and to cut syringe tips manually using the appliance |
Type | Portable and electrical |
Principle Of Operation | Electro-melting type |
Housing/enclosure Shall Be Moulded Type And Shock Proof And Made Of Materials Such As ABS Plastics/glass Filled Nylone. | SS frame |
Suitable Sliding Tray Shall Be Provided For Collection Of Residue And Destroyed Syringe Tips. | Yes |
Provision To Burn The Needles And To Cut The Syringe Tips Shall Be Provided In The Unit. | Yes |
Comply In General To IS:302Pt-1/1979 To Ensure Safe And Reliable Operation In Normal Use. | Yes |
Manual Cutter Provided In The Equipment | Made of hardened blade of anti-corrossive SS material |
Rated Power In Watts | 60 |
Sizes Of Injection Needles Of All Kinds Which Can Be Destroyed | Dia ranging from 0.4mm to 1.6mm(26SWG to 14SWG) with length 12.5mm to 80mm |
Powe ON/OFF Switch With Power "ON"indicator Shall Be Provided | Yes |
Shall Be Provided With Fuse And Power Cord Of Min 3m Length And Earthing Point | Yes |
Power Supply | 207V to 253V,50Hz AC supply |
Quantity(Nos) Of Needles Of 1mm Dia And 80mm Length That Can Be Destroyed In Continuous Operation Of 5 Minutes . | 1 |
Occupational Health And Safety Certification From Approved Bodies As Per OHSAS 18001-2007 | Yes |
Quality Management System For Medical Devices From Approved Bodies As Per ISO 13485-2009 | Yes |
CE Certification From Approved Bodies | Yes |
Performance Tests Shall Consist Of Capacity Test As Per Clause 14 Above ,power Consumption Test At Maximum Load And Ability To Cut Syringe Tips | Yes |
Test Report No. & Date | NA |
Name Of Lab | NA |
Address Of Lab | NA |
Manual | To be supplied by the supplier |
Safety Tests Shall Consist Of Marking,protection Against Electric Shock,temperature Rise ,IR,Leakage Current ,HV,Stability And Mechanical Hazards,mechanical Strength,constructional Requirements,creepage Distance And Clearances,resistance To Heat,fire And Tracking As Per IS:302Pt-1/79 | NA |
Availabiity Of The Test Report From Central Govt./NABL Approvel/ ILAC Acredited Lab Covering Performance Tests And Safety Tests | Yes |