A compact UV-Vis spectrophotometer featuring an integrated Fluorescence module. It can be used to increase the measuring range for biomolecules like nucleic acids and proteins by a factor of 1000, compared to UV measurements. Automatic evaluation and storage of measurement data with clear presentation of results. Data transfer via USB interface. Display - VGA TFT Display 5.7 inch, Monochromator - Holographic aberration-corrected concave grating, Wavelength Range user selectable - 200nm to 830nm, Fluorescence excitation 470 nm and emission 520,560 nm, Wavelength Resolution - 0.5 nm, Wavelength Accuracy - /-1nm, Repeatability - 0.5 nm, Spectral Bandwidth - 4nm, Type of Detector - CMOS Photodiode Array, Measuring Modes - Absorbance, concentration, scan absorbance wavelength spectrum, Operating Mode - Pre-programmed and freely programmable, Light Source - Xenon flash lamp. Eppendorf microCuvette G1.0 is a Microvolume measuring cell for measuring 1.5-10 microLiter sample volumes..
Other Specifications Download Specification | |
---|---|
Spectrophotometer Type | Microprocessor Based |
Display | High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC |
Monochromator | Concave holographic grating with 1200 lines/mm |
Wavelength Range (user Selectable) | 340-960 nm |
Wavelength Resolution | 0.2 nm |
Wavelength Accuracy | +-1 nm |
Repeatability | +-1 nm |
Spectral Bandwidth | < 2 |
Type Of Detector | Photodiode |
Measuring Modes | Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards. |
Operating Mode | Single wavelength: Measuring 1, ABS and CONC |
Light Source | Tungstan Halogen Lamp |
Power | 230 V +- 10 50Hz |
Standard Accessories To Be Supplied | 10 nm path length four (4 nos.) Cuvettes matched with 3 percent |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | EN 61010-1 UL 61010-1, CAN/CSA C22.2 No. 61010-1 |
Power Requirements | 230 V +- 10, 50 Hz |
Operating Temperature | 15 to 35 Degree C |
Operating Humidity (RH) | 25 to 70 % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | 2014/30/EU |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | Yes |
Test Report No. | 20110826-E158089 |
Test Report Date | 26-08-2011 |
Name Of The Lab | Underwriters Laboratories |
Warranty | 1 Year |