Keysight N9938A- 26.5GHz Handheld Spectrum Analyzer, battery Operated, Rugged and weather resistant.Meets MIL-PRF-28800F Class 2, no vents or fans, wide operating temp of minus 10 to 55 degree C, Meets IEC/EN 60529 IP53 requirements, Type tested and meets Procedure 1 requirements for operation in explosive environments. MIL-810G, Method 511.5 Portable Compact package Provides best amplitude accuracy plus/minus 0.5 dB, no warm up required Future Upgradable options at additional cost such as- Real-time spectrum analyzer RTSA with 10 MHz bandwidth and minimum detectable signal as short as 22ns, Full-band tracking generator, Built-in power meter, Pulse Measurement, Interference Analyzer, Channel Scanner, Pre-amplifier, Built-in GPS and frequency counter, Analog demodulation. Standard 3-Year Warranty. Refer OEM datasheet for detail specification. Standard delivery period is 6 weeks. Prices indicated included GST@18percent..
Other Specifications Download Specification | |
---|---|
Description Of The Item | Spectrum analyzers 26.5GHz |
Frequency Range | 9 kHz to 26.5 GHz |
Aging Rate Of Spectrum Analyzer Frequency: Better Than 1x10-6 /Year | Yes |
Frequency Counter Resolution(min.) | 1 Hz |
Frequency Span | 0Hz(zero span),100Hz to 26.5GHz |
Spectral Purity:SSB Phase Noise @10KHz Offset Shall Be Less Than -90dBc/Hz At 1GHz | Yes |
Spectral Purity :Harmonics @ -40 DBm Level Shall Be Less Than Or Equal To -60dBc | Yes |
Sweep Time @span >or= 1 KHz | NA |
Sweep Time @span = 0 Hz | 200 micro S to 20 S |
Resolution Bandwidth (-3dB) | 10 Hz to 3 MHz in steps |
Video Bandwidth | 10Hz to 1 MHz in steps |
Amplitude : Max Input Level For Protection | +27 dBm |
Amplitude : Max Input Level For Measurement | +20 dBm |
Amplitude :level Measurement Uncertainity | |
Average Noise Level Displayed | 300HzRBW,10Hz VBW: |
Markers | Normal and delta |
Trigger | Free run, video, External |
Display | TFT Colour display with provision for three traces |
Interface | USB and LAN |
Measurement Features | CN ratio,AM,time domain power measurement,channel power measurement,Phase noise measurement,TOI,OBW |
Software For Transfer Of Screen Shot/measurement Data To PC | Yes |
2-port Vector Network Analyzer Functionality ,for General Purpose Testing Applications In The Field,specially For Cable And Antenna Measurements, Is Available. | NA |
Frequency Range For 2port Vector Network Analyzer | 2 MHz to 3 GHz |
Frequency Resolution (CW ON) For 2port Vector Network Analyzer | NA Hz |
Return Loss For 2port Vector Network Analyzer | NA dB |
VSWR For 2port Vector Network Analyzer | NA |
Cable Loss For 2port Vector Network Analyzer | NA dB |
For 2port Vector Network Analyzer: Calibration With Precision Calibration Components-open, Short, Load Should Be Included To Achieve Directivity > 35 DB | NA |
For 2port Vector Network Analyzer: Distance-to-fault In Co-axial Cables Shall Be Possible | NA |
Phase Range For 2port Vector Network Analyzer | NA degree |
Phase Resolution For 2port Vector Network Analyzer | NA deg |
RF Output Port For 2port Vector Network Analyzer | Yes |
Dynamic Range For 2port Vector Network Analyzer | NA dB |
Nominal Port Power Output For 2port Vector Network Analyzer:Hi Power/Lo-Power Selectable(power Levels Corresponding To Hi And Lo Levels Shall Be Indicated By Bidder. | NA |
Measurement Of S11 & S21 And Smith Chart Measurement Shall Be Possible | NA |
For 2port Vector Network Analyzer :Vector Voltmeter For Comparision And Verification Of The Phase Of The Phase-matched Cables Shall Be Available | NA |
For 2port Vector Network Analyzer:Group Delay Measurement Shall Be Possible.Details To Be Declared | NA |
Operating Temperature | 5 to 40 degree C & 80 RH 40 degree C |
Power Requirement | 230 +- 10 VAC, 50Hz & External DC Input +12.5V to 15V with rechargeable battery of 1500 mAH or higher |
Warranty | 3 years |
OEM Shall Have ISO:14001 Certification For Their Manufacturing Works | yes |
One Set Of Operation And Maintenance Manual Shall Be Supplied | yes |
Standard Accessories Shall Be Supplied | yes |
Availability Of Test Reports From Central Govt/NABL/ILAC Accredited Lab Covering All The Features & Functional Requirements Including Environmental Tests As Per Specifications | yes |
Test Report No | NTH(SR)/EL(C )/2014/465C |
Test Report Date | 28-11-2014 |
Name Of Lab | National Test House (SR) |
Address Of The Lab | NTH Taramani, Chennai |