The Eppendorf BioPhotometer D30, Single-beam absorption photometer with reference beam, is an established standard in the life sciences. Quantification of Nucleic acids, Proteins direct and colorimetric, Bacteria growth measurement OD 600. Automatic evaluation and storage of measurement data with clear presentation of results. Data transfer via USB interface. Routine pre-defined wavelengths are easily selectable, allowing for quick and simple programming. Light Source - Xenon flash lamp. Type of Detector - CMOS Photodiode. Fixed wavelength in nm - 230, 260, 280, 320, 340, 405, 490, 562, 595, 600. Data output via JPEG screenshots, Microsoft Excel or secure PDF files to a standard USB flash drive or via email. Compatible with semi-micro and micro-volume measuring cells as well as standard cuvettes..
Other Specifications Download Specification | |
---|---|
Spectrophotometer Type | Microprocessor Based |
Display | High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC |
Monochromator | Concave holographic grating with 1200 lines/mm |
Wavelength Range (user Selectable) | 340-960 nm |
Wavelength Resolution | 0.2 nm |
Wavelength Accuracy | +-1 nm |
Repeatability | +-1 nm |
Spectral Bandwidth | < 2 |
Type Of Detector | Photodiode |
Measuring Modes | Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards. |
Operating Mode | Single wavelength: Measuring 1, ABS and CONC |
Light Source | Tungstan Halogen Lamp |
Power | 230 V +- 10 50Hz |
Standard Accessories To Be Supplied | 10 nm path length four (4 nos.) Cuvettes matched with 3 percent |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | EN 61010-1 UL 61010-1, CAN/CSA C22.2 No. 61010-1 |
Power Requirements | 230 V +- 10, 50 Hz |
Operating Temperature | 15 to 35 Degree C |
Operating Humidity (RH) | 25 to 70 % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | 2014/30/EU |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | Yes |
Test Report No. | 20110826-E158089 |
Test Report Date | 26-08-2011 |
Name Of The Lab | Underwriters Laboratories |
Warranty | 1 Year |