Keysight N9918A offered configuration has full 2-port Vector Network Analyzer from 30KHz to 26.5GHz and Spectrum Analyzer from 5KHz to 26.5GHz. 100dB Dynamic Range, 42dB Directivity, plus-minus 0.7ppm frequency accuracy, built in CalReady, Data Points upto 10,001, user interface and Smith chart and polar display, Weather resistant, field-friendly design, and no fans or vents with full MIL-PRF-28800F Class 2 compliance, Expand the capabilities with optional built-in power meter, vector voltmeter, GPS Receiver, Channels Scanner, Independent signal generator, Interference Analyzer, Pulse Measurement, Built-in Variable DC Source, TDR Measurement, ECal Support. USB, LAN Connectivity, 3.5 hour battery operating time, Standard 3 Year Warranty. Offered configuration is N9918A with option 010, 210, 211, 233, 235, 302, 330, 350, N9910X-708, 880 and 8552D. Refer Keysight datasheet for detailed specifications. Prices indicated including GST @18 percent. Delivery Period 60 Days..
Other Specifications Download Specification | |
---|---|
Description Of The Item | Two port vector network analyzers 6GHz |
Frequency Range | 9 kHz to 6 GHz |
Aging Rate Of Spectrum Analyzer Frequency: Better Than 1x10-6 /Year | Yes |
Frequency Counter Resolution(min.) | 1 Hz |
Frequency Span | 0Hz(zero span),100Hz to 26.5GHz |
Spectral Purity:SSB Phase Noise @10KHz Offset Shall Be Less Than -90dBc/Hz At 1GHz | Yes |
Spectral Purity :Harmonics @ -40 DBm Level Shall Be Less Than Or Equal To -60dBc | Yes |
Sweep Time @span >or= 1 KHz | NA |
Sweep Time @span = 0 Hz | 100 micro S to 20 S |
Resolution Bandwidth (-3dB) | 10 Hz to 3 MHz in steps |
Video Bandwidth | 10Hz to 1 MHz in steps |
Amplitude : Max Input Level For Protection | +27 dBm |
Amplitude : Max Input Level For Measurement | +20 dBm |
Amplitude :level Measurement Uncertainity | |
Average Noise Level Displayed | 300HzRBW,10Hz VBW: |
Markers | Normal and delta |
Trigger | Free run, video, External |
Display | TFT Colour display with provision for three traces |
Interface | USB and LAN |
Measurement Features | CN ratio,AM,time domain power measurement,channel power measurement,Phase noise measurement,TOI,OBW |
Software For Transfer Of Screen Shot/measurement Data To PC | Yes |
2-port Vector Network Analyzer Functionality ,for General Purpose Testing Applications In The Field,specially For Cable And Antenna Measurements, Is Available. | Yes |
Frequency Range For 2port Vector Network Analyzer | 2 MHz to 6 GHz |
Frequency Resolution (CW ON) For 2port Vector Network Analyzer | 10 Hz |
Return Loss For 2port Vector Network Analyzer | 0.00 to 60 dB |
VSWR For 2port Vector Network Analyzer | 1.01 to 10.00 |
Cable Loss For 2port Vector Network Analyzer | 0.00 to 30.00 dB |
For 2port Vector Network Analyzer: Calibration With Precision Calibration Components-open, Short, Load Should Be Included To Achieve Directivity > 35 DB | Yes |
For 2port Vector Network Analyzer: Distance-to-fault In Co-axial Cables Shall Be Possible | Yes |
Phase Range For 2port Vector Network Analyzer | -180 to +180 degree |
Phase Resolution For 2port Vector Network Analyzer | 0.01 deg |
RF Output Port For 2port Vector Network Analyzer | Yes |
Dynamic Range For 2port Vector Network Analyzer | 60 dB |
Nominal Port Power Output For 2port Vector Network Analyzer:Hi Power/Lo-Power Selectable(power Levels Corresponding To Hi And Lo Levels Shall Be Indicated By Bidder. | Selectable and Hi-power= + 3 dB at 3 GHz Low-power = -45dBm |
Measurement Of S11 & S21 And Smith Chart Measurement Shall Be Possible | Yes |
For 2port Vector Network Analyzer :Vector Voltmeter For Comparision And Verification Of The Phase Of The Phase-matched Cables Shall Be Available | Yes |
For 2port Vector Network Analyzer:Group Delay Measurement Shall Be Possible.Details To Be Declared | Yes Available |
Operating Temperature | 5 to 40 degree C & 80 RH 40 degree C |
Power Requirement | 230 +- 10 VAC, 50Hz & External DC Input +12.5V to 15V with rechargeable battery of 1500 mAH or higher |
Warranty | 3 years |
OEM Shall Have ISO:14001 Certification For Their Manufacturing Works | yes |
One Set Of Operation And Maintenance Manual Shall Be Supplied | yes |
Standard Accessories Shall Be Supplied | yes |
Availability Of Test Reports From Central Govt/NABL/ILAC Accredited Lab Covering All The Features & Functional Requirements Including Environmental Tests As Per Specifications | yes |
Test Report No | NTH(SR)/EL(C )/2014/465G |
Test Report Date | 11/28/2014 |
Name Of Lab | National Test House |
Address Of The Lab | National Test House, Taramani, Chennai |