Needle and syringe destroyer able to burn needle electrically and Cutting of hub mechanically.CUTTING BLADE Material Stainless steel S.S.304,Thickness 16G 0.5 mm with one side sharpen,Number of cuts of 10,000 cuts.ABS Plastic Body Made of Body Virgin ABS or NAYLON GLASS FILL Molded Plastic.SPRING Made of steel wire for mechanical spring.CAPACITY RANGE Needles from 14-28 gauge diameter and from 10-100 mm length - and Fitted with all types of needle fittings, egg, fixed lure lock, lure ship of Snap-on.Filling Container shall accept 200 cut needles along with the hub when it is 3/4 full.Transformer IS Copper 10V-10Amp. Needle Burning Point 18G Electric grade Copper Plate.Mains Cord Wire 3pin Copper with Grommet 9FT ISI. Fuse 2amp ISI.electrically operated needles and syringes destroyer 100w is automated thermostate control..
Other Specifications Download Specification | |
---|---|
Purpose | To destroy the used injection needles by instant electromelting process and to cut syringe tips manually using the appliance |
Type | Portable and electrical |
Principle Of Operation | Electro-melting type |
Housing/enclosure Shall Be Moulded Type And Shock Proof And Made Of Materials Such As ABS Plastics/glass Filled Nylone. | MOULDED TYPE,SHOCK PROOF,ABS |
Suitable Sliding Tray Shall Be Provided For Collection Of Residue And Destroyed Syringe Tips. | Yes |
Provision To Burn The Needles And To Cut The Syringe Tips Shall Be Provided In The Unit. | Yes |
Comply In General To IS:302Pt-1/1979 To Ensure Safe And Reliable Operation In Normal Use. | Yes |
Manual Cutter Provided In The Equipment | Made of hardened blade of anti-corrossive SS material |
Rated Power In Watts | 100 |
Sizes Of Injection Needles Of All Kinds Which Can Be Destroyed | Dia ranging from 0.4mm to 1.0mm(26SWG to 18SWG) with length 12.5mm to 80mm |
Powe ON/OFF Switch With Power "ON"indicator Shall Be Provided | Yes |
Shall Be Provided With Fuse And Power Cord Of Min 3m Length And Earthing Point | Yes |
Power Supply | 207V to 253V,50Hz AC supply |
Quantity(Nos) Of Needles Of 1mm Dia And 80mm Length That Can Be Destroyed In Continuous Operation Of 5 Minutes . | 30 |
Occupational Health And Safety Certification From Approved Bodies As Per OHSAS 18001-2007 | Yes |
Quality Management System For Medical Devices From Approved Bodies As Per ISO 13485-2009 | Yes |
CE Certification From Approved Bodies | Yes |
Performance Tests Shall Consist Of Capacity Test As Per Clause 14 Above ,power Consumption Test At Maximum Load And Ability To Cut Syringe Tips | Yes |
Test Report No. & Date | EQP/G0344/1805/2015,05/06/2015 |
Name Of Lab | ELECTRONICS & QUALITY DEVELOPMENT CENTRE |
Address Of Lab | B/177/178,GIDC,ELECTRONICS ESTATE,SEC-25,GANDHINAGAR,GUJARAT |
Manual | To be supplied by the supplier |
Safety Tests Shall Consist Of Marking,protection Against Electric Shock,temperature Rise ,IR,Leakage Current ,HV,Stability And Mechanical Hazards,mechanical Strength,constructional Requirements,creepage Distance And Clearances,resistance To Heat,fire And Tracking As Per IS:302Pt-1/79 | YES |
Availabiity Of The Test Report From Central Govt./NABL Approvel/ ILAC Acredited Lab Covering Performance Tests And Safety Tests | Yes |