It is a double walled unit with stainless steel interior and exterior sheet duly enamel painted. The pyramidicallid made of S.S. prevents falling of condensed water droplets on the specimens. Heating is achieved throughan immersion type heater and the temperature is controlled byIMPORTED THERMOSTAT hydraulic type thermostat with an accuracy of 0.5C. Thebath has an oscillating tray riding on ball rollers oscillated through a geared variable speedmotor. Shaking speed range is between 40 to 140 cycles per minute and shaking tray canhold test tubes or Erlenmeyer flasks of 25 ml or 50 ml or 100 ml or 250 ml or 500 ml..
Other Specifications Download Specification | |
---|---|
Purpose Of Waterbath. | To test clinical thermometer (Liquid in glass(LIG) &electrical) with maximum device within accuracy of ± 0.15 C |
Minimum Numbers Of Thermometers Which Can Be Simultaniously Calibrated In Water Bath. | 20 Nos |
Reference Thermometer Confirming To IS:4825/1982 | YES |
Min. Resolution Of Reference Thermometer | 0.01 °C |
Min.uncertainty Of Reference Thermometer(at Coverage Factor K=2) | 0.02 °C |
The Comparator Water Bath Either Screw Type Or Bubble Type As Per IS:6274/1971,should Be Constant And Uniform Within ±0.01 °C,stability ±0.01 °C And Should Have Digital Indication With Resolution Of 0.01 °C. Thermocouple Shall Be Calibrated And Shall Have Calibration Certificate. | Yes |
Mechanism To View The LIG Thermometer With Magnification Of 10X. | Yes |
Mechanism To Hold Thermometer In Water For At Least 20 Seconds. | Yes |
Centrifuge To Drive Mercury Level Well Below 25C At 600 Rpm. | Yes |
Thermometer Calibrator Shall Be Capable To Calibrate The LIG Thermometer As Per IS:12622/1989 | Yes |
All The Gazette Of India Related To Legal Metrology Issued Time To Time Are Applicable To The Specification As Well As To The Product.Also As Per The Gazette Of India Extraodinary,The Legal Metrology Act 2009,cl.22, Approval Of Model Before Manufacturing And Cl.23.Prohibition On Manufacture,repair Or Sale Without Licence For Product Shall Be Applicable On Manufacture Of Measuring Instrument. | Yes |
Availability Of Calibration Test Reports Of The Reference Thermometer From NPL Or Central Govt/NABL/ILAC Accredited Lab . | Yes |
Availability Of Type Test Reports Of The Product From Central Govt/NABL/ILAC Accredited Lab Showing Compliance To The Specification | Yes |
Test Reports No. | N/A |
Test Reports Date | N/A |
Name Of The Labs | N/A |
Address Of The Lab | N/A |
Product Code Available On Supplier Website | Yes |
Product Code No | SDS0068 |