× All the transactions performed on 2.0 is available here. To access Market click "Home".

Insulation tester (Rishabh)(RISH Insu 20)

` 5,410.00

  • MRP: `6020.36

Insulation Tester.

  • Availability: 94 In Stock
  • Min. Qty. Per Consignee: 1
  • Measurement Range - 0.01 M Ohm to 10 G Ohm 0.01 M Ohm to 10 G Ohm 0.01 M Ohm to 10 G Ohm
  • Test Voltage - 50, 100, 250, 500 and 1000 Volt
  • AC/DC Voltage Measurement - 600 Volt
  • AC/DC Voltage Measurement Accuracy - 2 %
  • Pass Fail Function - Yes
Other Specifications Download Specification
Type Hand Held
Measurement Range 0.01 M Ohm to 10 G Ohm 0.01 M Ohm to 10 G Ohm 0.01 M Ohm to 10 G Ohm
Test Voltage 50, 100, 250, 500 and 1000 Volt
AC/DC Voltage Measurement 600 Volt
AC/DC Voltage Measurement Resolution 0. 1 %
AC/DC Voltage Measurement Accuracy 2 %
Auto Discharge Time Less than 0.5 Second
Live Circuit Detection Inhibit test terminal voltage >30 V prior to initialization of test
Pass Fail Function Yes
Data Logging NA
Software Not Available
Safety Rating Category III 600 V of IEC 61010
Leakage Current 2 milliamper
Operating Temperature Range -20 to +40 Degree C
Operating Humidity (RH) 90 %
Power Requirements Battery Operated
Type Of Battery 6 X 1.5V cells IEC LR6
PC Interface Available
Standard Accessories Supplied With The Equipment 1 Meter, 1 Cable set, Round Pencil cell-6nos, 1 Spare fuse, 1 Belt, 1 User manual, 1 Warranty Card, 1 Test Certificate, 1 Crocodile Clip set, 1 Safety Cover
Dimensions (H X W X L) 195 X 84 X 35 mm x mm x
Weight 500g gms
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification Including Environmental Tests As Under : (a) Dry Heat : For 16 Hrs. At A Temp. Of 55 Degree C In Accordance With IS:9000/part-3/section-5/1977 (reaffirmed In 2007). (b) Cold Test : For 4 Hrs. At A Temp. Of -20 Degree C In Accordance With IS:9000/part-2/section-4/1977 (reaffirmed In 2007). (c) Damp Heat (Cyclic)Test : For 2 Cycles Of 24 H Each At A Temp. Of 40 Degree C & 95% RH In Accordance With IS:9000/part-5/section-1/ 1991 (reaffirmed In 2007). After Environmental Conditioning Sequence As Above And A Recovery Period Of Two Hours, The Instrument Shall Be Checked For Complete Functional Parameters, Which Should Be With In Limits.(Test Reports Are To Be Furnished When Demanded By Buyer) Yes
Test Report No. 375
Test Report Date 13/09/2016
Name Of The Lab NABL approved
Warranty 1