GENESYS 10S UV-Vis, Europlug UK line cords Part No. 840-209700 Specifications-- Optical Design Dual-Beam-- Spectral Bandwidth 1.8 nm-- Light Source Typical Lifetime Xenon Flash Lamp 5 years, 3 years guaranteed-- Detectors Dual Silicon Photodiodes-- Wavelength Range 190-1100 nm-- Accuracy 1.0 nm-- Repeatability 0.5 nm-- Slew Speed 11,000 nm/min-- Scan Speed Up to 3600 nm/min-- Data Interval for Scanning 0.1, 0.2, 0.5, 1.0, 2.0, 5.0 nm-- Photometric Linear Range Up to 3.5 A at 260 nm-- Display -0.1 A-- -1.5-125T-- 9999 C-- Accuracy 0.005 A at 1.0 A 0.010 A K2Cr2O7-- Noise 0.00025 at 0.0 A-- 0.00050 at 1.0 A-- 0.00080 at 2.0 A-- RMS at 260 nm-- Drift 0.0005 A/hr-- Stray Light 0.08T at 220, 340 nm NaI, NaNO2-- 1.0T 198-200 nm KCl-- Display Graphical with LCD backlight-- 9.7 7.1 cm 3.8 2.8 in-- Power Requirements Selected Automatically 100-240 V-- 50-60 Hz--.
Other Specifications Download Specification | |
---|---|
Spectrophotometer Type | Microprocessor Based |
Display | High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC |
Monochromator | Concave holographic grating with 1200 lines/mm |
Wavelength Range (user Selectable) | 340-960 nm |
Wavelength Resolution | 0.2 nm |
Wavelength Accuracy | +-1 nm |
Repeatability | +-1 nm |
Spectral Bandwidth | |
Type Of Detector | Photodiode |
Measuring Modes | Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards. |
Operating Mode | Single wavelength: Measuring 1, ABS and CONC |
Light Source | Tungstan Halogen Lamp |
Power | 230 V +- 10 50Hz |
Standard Accessories To Be Supplied | 10 nm path length four (4 nos.) Cuvettes matched with 3 percent |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | CE Certificate On demand |
Power Requirements | 230 V +- 10, 50 Hz |
Operating Temperature | Ambient Degree C |
Operating Humidity (RH) | Ambient % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | NA |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | NA |
Test Report No. | 0 |
Test Report Date | 0 |
Name Of The Lab | 0 |
Warranty | 1 Year |