TABLE TOP PULSE OXIMETER BEDSIDE MODEL, 7 SEGMENT LED FOR SP2,PULSE RATE AND FOR PI.2.5 LCD FOR SPO2,GRAPH PIETHYSMOGRAPH. WITH RESISTANCE TO SURROUNDING LIGHT.THE DEVIATION BETWEEN THE VALUE MEASURED IN THE CONDITIONS OF MAN MADE LIGHT OR INDOOR NATURAL LIGHT AND THAT OF DARK ROOM IS LESS THAN 1 PERCENT.THE INTERNAL POWER SUPPLY IS 3.6V DC TO 4.2V DC.THE BATTERY BACK UP IS APPROX 3 TO 4 HOURS...
Other Specifications Download Specification | |
---|---|
GMDN Name And Code | Pulse oximeter CT 1446 |
"Pulse Oxymeter, Line Powered:A Mains Electricity (AC-powered) Photoelectric Device Intended For The Continuous Transcutaneous Measurement And Display Of Haemoglobin Oxygen Saturation (SpO2). The Signals, Typically Produced By Light-emitting Diodes (LEDs) And A Receiving Detector In A Probe, Or Directly Built-in, Are Used To Make The Measurements Using The Principle Of Spectrophotometry. The Oximeter Displays The SpO2 Values And May Calculate / Display Other Parameters, E.g., Pulse Rate, Electrocardiogram (ECG). The Device Is Typically Used At The Bedside." | yes |
SpO2 Measurement Range At Least 40-70 And 70 To 99 %, Minimum Gradation 1% | yes |
Accuracy Of SpO2 Better Than ± 1% For Range 40-70 And Better Than ± 3% For Range 70-99. | yes |
Pulse Rate Range At Least 30 To 240 Bpm, Minimum Gradation 1 Bpm | yes |
Accuracy Of Pulse Rate Better Than ± 5 Bpm | yes |
Signal Strength Or Quality To Be Visually Displayed | yes |
Audiovisual Alarms Required: High And Low SpO2 And Pulse Rate (operator Variable Settings), Sensor Disconnected,sensor Failure, Low Battery | yes |
Display Type | TFT Screen |
Memory | Minimum 24 hrs trend memory for SpO2 & PR |
User's Interface | Easily accessible touch button to operate the machine |
Inbuilt Software And/or Standard Of Communication | yes |
Weight | 1600 gram |
Mobility, Portability | Protective splashproof case for clean storage and safe transport |
Display Must Allow Easy Viewing In All Ambient Light Levels | yes |
Noise Lavel | 38 dba |
Heat Dissipiation: Should Maitain Nominal Temp And The Heat Should Be Disbursed Through A Exhaust Cooling Fan | yes |
Power Input And Frequency | 220 to 240V, 50 Hz |
Power Consumption | 220 to 240v,50Hz Watt |
Power Cable Length | 3 meter |
Electrical Protection Provided By Fuses In Both Live And Neutral Supply Lines. | yes |
"Operating Condition:Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances. Circumstances" | yes |
"STANDARDS AND SAFETY:Should Be FDA / CE Approved Product ISO 80601-2-61-2011: Medical Electrical Equipment- Part 2-61: Particular Requirements For The Basic Safety Snd Essential Performance Of Pulse Oxymeter. Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1, EMC Safety Confirms To IEC 60601-1-2 Standard Requirement; Manufacturer / Supplier Should Have ISO 13485 Certificate For Quality Standard Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (essential Performance Of Electrocardiographs)" | yes |
Warranty | 3 years warranty & one year with min. 3 free servicing. |
Others: List Of Important Spare Parts And Accessories With Their Part Number And Costing. Certificate Of Callibration And Inspection To Be Provided | yes |