12-channel machines, most advanced ECG machines till date,simultaneously acquire and display all 12 leads at any time. BPLs offerings in 12-channel ECGs include Cardiart 9108, Cardioline ar2100view and Cardioline 200, equiped with Glasgow Algorithm.
Other Specifications Download Specification | |
---|---|
GMDN Name And Code | Multichannel Electrocardiographic , CT-1115 |
Functional Requirements: Electricity (AC-powered) Bedside Device Designed To Continuously Detect, Measure, And Display A Patient's Electrocardiogram (ECG) Through Leads And Sensors Attached To The Patient, It Also Typically Display Heart Rate. The Device Is Typically Equipped With Audible And/or Visual Alarms That Are Triggered When The Patient's Parameters Drop Below Or Exceed Pre-set Limits. Continuous Display Of Patient ECG And Heart Rate On Screen,Allows Display Of Single, 5 Lead ECG Or Simultaneous Display Of At Least 5 Waves Selected From Up To 12 Points.Operator Can Set Audiovisual Alarm Levels For Low Or High Heart Rate.Operates From Mains Voltage Or From Internal Rechargeable Battery. | yes |
Heart Rate Measurement Range To Be At Least 30 To 250 Bpm, With Accuracy Better Than ± 5 Bpm | yes |
Heart Rate Trend Display Of At Least Previous 24 Hours | yes |
Arrhythmia Detection Facility Required; Minimum Gradation Of 1 Bpm | yes |
Audiovisual Alarms Required: High And Low Heart Rate (operator Variable Settings), Cardiac Arrhythmia, Sensor / Wire Disconnected, Low Battery | yes |