The PinAAcle500 is a standalone flame-only atomic absorption AA spectrophotometer featuring a fully integrated operational benchtop AA system with a simplified touch screen user interface. Monochromator- Littrow design with motorized drive for automatic Wavelength selection and peaking. Wavelength range 180 900 nm, Diffraction grating 1800 lines/mm blazed at 236 nm and 597 nm, Grating area 64 x 72 mm Reciprocal linear dispersion 1.6 nm/ mm nominal Focal length 267 mm. Spectral Bandwidths User selectable automatic slit widths of 2.0, 0.7 and 2.0 nm at their optimized slit height. Detector- Wide range segmented solid-state detector, including a built- in low noise CMOS charge amplifier array. Built-in continuum source double-beam background correction using a high intensity deuterium arc lamp. All the PCBAs are conformal coated for complete resistance to corrosion..
Other Specifications Download Specification | |
---|---|
Computer Controlled True Double Beam Atomic Absorption Spectrophotometer | Yes |
Eight Lamp Automatic Turret With Independent Eight Power Supplies With Coded Lamp Compatible | Yes |
Monochromator With Wavelength Range 190-900 Nm. | Yes |
Holographic Grating Of High Density Of Not Less Than 1800 Lines/mm | Yes |
System Have Wavelength Locating By Automatic Peak Searching, Auto Loading Of All Parameters And Auto Bandwidth Selection And Continuously Variable Selection 0.2-2.0 Nm With D2 Lamp Background Correction | Yes |
Titanium Burners For Acetylene Nitrous Oxide (C2H2- N2O) & Acytelene-Air (C2H2-AIR) With Precise Knobs For Burner Optimization I.e. Height, Rotational And Lateral | Yes |
Fully Inert Nebulizer | Yes |
Gas System With Automatic Flame Changeover, Full Safety Interlock Including Pressure Sensors On Both Line, Power Failure Protection, Burner Interlock And Flame Sensor, Flame Ignition Should Be Automatic | Yes |
Burner Movement Thorough Computer, All Three Directional, Height, Rotation Lateral | Yes |
PTFE Spray Chamber And Adjustable Impact Bead Aerosol | Yes |
Single Element Coded Hollow Cathode Lamps : Cu, Fe, Zn, Mn, Mg B, K, Mo. | Yes |
System Have Facilities Of Repeat Of Result Of Same Sample And Date Treatment. Automatic Calculation Of Percentage In Base Material | Yes |
Standard Accessories To Be Supplied : 1. Acetylene And Nitrous Oxide Gas Cylinders Should Have Explosive Certification With Regulator 2. Two Filled Cylinders For Each Gas, 3. Air Compressor, 4. Air Filter, 5. Spares And Consumables For 2 Years Operations. 6. SS Exhaust Fume Hood With Inert Centrifugal Blower, 7. Instruction Manual 8. Circuit Diagram | Yes |
Spare One Set Hollow Cathode Lamps Of Cu, Fe, Zn, Mn. | Yes |
100 Ml Standard Solutions Of 1000 Ppm For Each Element Under Guarantee Certificate | Yes |
. Training At The Time Of Installation Up To User Level Satisfaction | Yes |
PC Software | Available |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | Yes |
Power Requirements | 230 V +- 10, 50 Hz |
Operating Temperature | 0 to 40 Degree C Degree C |
Operating Humidity (RH) | 0 to 95% % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | Yes |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | Yes |
Test Report No. | EN 61010-1,CSA C22.2 No.1010 |
Test Report Date | NA |
Name Of The Lab | NA |
On-site Warranty | 1 Year |
Availability & Supply Of Spare Parts & Accessories | For next 10 years |