ECG machine 12 channel- Functional requirements electricity AC-powered bedside device designed to continuously detect, measure, and display a patients electrocardiogram ECG through leads and sensors attached to the patient, it also typically display heart rate. The device is typically equipped with audible and/or visual alarms that are triggered when the patients parameters drop below or exceed pre-set limits. Continuous display of patient ECG and heart rate on screen,Allows display of single, 5 lead ECG or simultaneous display of at least 5 waves selected from up to 12 points.Operator can set audiovisual alarm levels for low or high heart rate.Operates from mains voltage or from internal rechargeable battery. - yesHeart rate measurement range to be at least 30 to 250 bpm, with accuracy better than 5 bpm - yesHeart rate trend display of at least previous 24 hours - yesAudiovisual alarms required high and low heart rate operator variable settings, cardiac arrhythmia, sen.
Other Specifications Download Specification | |
---|---|
GMDN Name And Code | Multichannel Electrocardiographic , CT-1115 |
Functional Requirements: Electricity (AC-powered) Bedside Device Designed To Continuously Detect, Measure, And Display A Patient's Electrocardiogram (ECG) Through Leads And Sensors Attached To The Patient, It Also Typically Display Heart Rate. The Device Is Typically Equipped With Audible And/or Visual Alarms That Are Triggered When The Patient's Parameters Drop Below Or Exceed Pre-set Limits. Continuous Display Of Patient ECG And Heart Rate On Screen,Allows Display Of Single, 5 Lead ECG Or Simultaneous Display Of At Least 5 Waves Selected From Up To 12 Points.Operator Can Set Audiovisual Alarm Levels For Low Or High Heart Rate.Operates From Mains Voltage Or From Internal Rechargeable Battery. | yes |
Heart Rate Measurement Range To Be At Least 30 To 250 Bpm, With Accuracy Better Than ± 5 Bpm | yes |
Heart Rate Trend Display Of At Least Previous 24 Hours | yes |
Arrhythmia Detection Facility Required; Minimum Gradation Of 1 Bpm | yes |
Audiovisual Alarms Required: High And Low Heart Rate (operator Variable Settings), Cardiac Arrhythmia, Sensor / Wire Disconnected, Low Battery | yes |
User's Interface | manual |
In Built Software And/or Standard Of Communication | yes |
Weight | 1500 gram |
Mobility, Portability | Protective splashproof case for clean storage and safe transport |
Display Must Allow Easy Viewing In All Ambient Light Levels | yes |
Noise Lavel | 45 dba |
Heat Dissipiation: Should Maitain Nominal Temp And The Heat Should Be Disbursed Through A Exhaust Cooling Fan | yes |
Power Input And Frequency | 220 to 240V, 50 Hz |
Power Consumption | 220 to 240V, 50hz watt |
Power Cable Length | 3 meter meter |
Electrical Protection Provided By Fuses In Both Live And Neutral Supply Lines. | yes |
"Operating Condition:Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances" | yes |
"STANDARDS AND SAFETY:FDA / CE Approved Product; Manufacturer / Supplier Should Have ISO 13485 Certificate For Quality Standard. Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (essential Performance Of Electrocardiographs)" | yes |
Warranty | 3 years warranty & one year with min. 3 free servicing. years |
Others: List Of Important Spare Parts And Accessories With Their Part Number And Costing. Certificate Of Callibration And Inspection To Be Provided | yes |