12 Channel ECG Machine with Display High resolution backlight 4.7 LCD screen Detailed 12 lead ECG Measurement Interpretation software Multiple print formats from integrated 12 channel thermal printer on A4 size thermal paper12 lead rhythm printout on single A4 page Alphanumeric keyboard with direct function keys for routine operation On-board inbuilt memory to store 40 ECG recordings Mains battery operated Battery capacity up to 300 ECG printouts or 4 hours of normal use Pacemaker detection OPTIONALSpirometry. The Specifications published on GEM site are of Heart Rate Monitor whereas ECG machine has features of an ECG machine. hence HEart Rate Trends Arrhythmias as asked cannot be viewed on screen but ECG machine has got inbuilt Interpretation software which reads ECG and interprets for help of physician.
Other Specifications Download Specification | |
---|---|
GMDN Name And Code | Multichannel Electrocardiographic , CT-1115 |
Functional Requirements: Electricity (AC-powered) Bedside Device Designed To Continuously Detect, Measure, And Display A Patient's Electrocardiogram (ECG) Through Leads And Sensors Attached To The Patient, It Also Typically Display Heart Rate. The Device Is Typically Equipped With Audible And/or Visual Alarms That Are Triggered When The Patient's Parameters Drop Below Or Exceed Pre-set Limits. Continuous Display Of Patient ECG And Heart Rate On Screen,Allows Display Of Single, 5 Lead ECG Or Simultaneous Display Of At Least 5 Waves Selected From Up To 12 Points.Operator Can Set Audiovisual Alarm Levels For Low Or High Heart Rate.Operates From Mains Voltage Or From Internal Rechargeable Battery. | yes |
Heart Rate Measurement Range To Be At Least 30 To 250 Bpm, With Accuracy Better Than ± 5 Bpm | yes |
Heart Rate Trend Display Of At Least Previous 24 Hours | yes |
Arrhythmia Detection Facility Required; Minimum Gradation Of 1 Bpm | yes |
Audiovisual Alarms Required: High And Low Heart Rate (operator Variable Settings), Cardiac Arrhythmia, Sensor / Wire Disconnected, Low Battery | yes |
User's Interface | manual |
In Built Software And/or Standard Of Communication | yes |
Weight | 4050 gram |
Mobility, Portability | Protective splashproof case for clean storage and safe transport |
Display Must Allow Easy Viewing In All Ambient Light Levels | yes |
Noise Lavel | 20 DB dba |
Heat Dissipiation: Should Maitain Nominal Temp And The Heat Should Be Disbursed Through A Exhaust Cooling Fan | yes |
Power Input And Frequency | 220 to 240V, 50 Hz |
Power Consumption | 28 va watt |
Power Cable Length | 3 meter meter |
Electrical Protection Provided By Fuses In Both Live And Neutral Supply Lines. | yes |
"Operating Condition:Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances" | yes |
"STANDARDS AND SAFETY:FDA / CE Approved Product; Manufacturer / Supplier Should Have ISO 13485 Certificate For Quality Standard. Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (essential Performance Of Electrocardiographs)" | yes |
Warranty | 3 years warranty & one year with min. 3 free servicing. years |
Others: List Of Important Spare Parts And Accessories With Their Part Number And Costing. Certificate Of Callibration And Inspection To Be Provided | yes |