× All the transactions performed on 2.0 is available here. To access Market click "Home".

ATOMIC ABSORPTION SPECTROPHOTOMETER AAS PLUS (MOTRAS)(MSIPL AAS PLUS)

` 1,633,500.00

  • MRP: `1815000.00

Fully Automatic Atomic Absorption Spectrophotometer, AAS is a microcomputer based Instrument for absorption measurement. It is used for quantitative elemental analysis mainlylike elements as Copper, Iron, Zinc, Lead , Arsenic, etc., by measuring the Absorbance of a sample atomized in aflame. The data processing and control of the instrument is done by Microcontroller. The Basic Spectrophotometer consist of Monochromator, Atomizer, Gas contolbox, Turret Assembly.The major features are stand alone operation Wi-Fi connectivity for remote controlled highly reliable double beam optics, 8 lamp turret, continuously variable auto Slit setting, Automatic peak, lamp current setting, gain/ EHT control Unit. Double beam with Deuterium background Ebert design with 333 mm focal length,1800 line/mm grating and 185-900 nm wavelength range. Automatic emission wavelength selection and peaking.Continuously variable slits with 0.2 to 2 nmspectral bandwidth. Automatic wavelength scanning. Selected wide r.

  • Availability: 10 In Stock
  • Min. Qty. Per Consignee: 1
  • Eight Lamp automatic turret with independent eight power supplies with coded lamp compatible - Yes
  • Monochromator with Wavelength range 190-900 nm. - Yes
  • Holographic grating of high density of not less than 1800 lines/mm - Yes
  • Titanium burners for Acetylene nitrous oxide (C2H2- N2O) & Acytelene-Air (C2H2-AIR) with precise Knobs for burner optimization i.e. height, rotational and lateral - Yes
  • Burner movement thorough computer, all three directional, height, rotation lateral - Yes
Other Specifications Download Specification
Computer Controlled True Double Beam Atomic Absorption Spectrophotometer Yes
Eight Lamp Automatic Turret With Independent Eight Power Supplies With Coded Lamp Compatible Yes
Monochromator With Wavelength Range 190-900 Nm. Yes
Holographic Grating Of High Density Of Not Less Than 1800 Lines/mm Yes
System Have Wavelength Locating By Automatic Peak Searching, Auto Loading Of All Parameters And Auto Bandwidth Selection And Continuously Variable Selection 0.2-2.0 Nm With D2 Lamp Background Correction Yes
Titanium Burners For Acetylene Nitrous Oxide (C2H2- N2O) & Acytelene-Air (C2H2-AIR) With Precise Knobs For Burner Optimization I.e. Height, Rotational And Lateral Yes
Fully Inert Nebulizer Yes
Gas System With Automatic Flame Changeover, Full Safety Interlock Including Pressure Sensors On Both Line, Power Failure Protection, Burner Interlock And Flame Sensor, Flame Ignition Should Be Automatic Yes
Burner Movement Thorough Computer, All Three Directional, Height, Rotation Lateral Yes
PTFE Spray Chamber And Adjustable Impact Bead Aerosol Yes
Single Element Coded Hollow Cathode Lamps : Cu, Fe, Zn, Mn, Mg B, K, Mo. Yes
System Have Facilities Of Repeat Of Result Of Same Sample And Date Treatment. Automatic Calculation Of Percentage In Base Material Yes
Standard Accessories To Be Supplied : 1. Acetylene And Nitrous Oxide Gas Cylinders Should Have Explosive Certification With Regulator 2. Two Filled Cylinders For Each Gas, 3. Air Compressor, 4. Air Filter, 5. Spares And Consumables For 2 Years Operations. 6. SS Exhaust Fume Hood With Inert Centrifugal Blower, 7. Instruction Manual 8. Circuit Diagram Yes
Spare One Set Hollow Cathode Lamps Of Cu, Fe, Zn, Mn. Yes
100 Ml Standard Solutions Of 1000 Ppm For Each Element Under Guarantee Certificate Yes
. Training At The Time Of Installation Up To User Level Satisfaction Yes
PC Software Available
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " yes
Power Requirements 230 V +- 10, 50 Hz
Operating Temperature 15-35 Degree C Degree C Degree C
Operating Humidity (RH) %% %
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" yes
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) Yes
Test Report No. STRC/CAL/1703040026
Test Report Date 06/03/17
Name Of The Lab SIGMA
On-site Warranty 1 Year
Availability & Supply Of Spare Parts & Accessories For next 10 years

Price Dekho Price Dekho Price List in india