This instrument is designed to measure Solar Power in the range in nanometers. The good Spectral Range, Orientation and Angular detection of meter allow users to conduct the most precise quantitative measurements of Solar Power Radiation Solar Power Measurement with Orientation and Tilt Angle, Easy measurement for rate of daylight penetration Auto change for measuring range Auto power off with disable functionInstantaneous display Ave/Min/Max values, 20 points memory, Socket of tripod mounting, Magnetic mount Backlit LCD and 4 digits triple display, Irradiance Display Range 01500 W/M2, Best Measurement Range of100-1250 W/m2 and Resolution 1W/m2 along with Temperature Display Range -30-125 Deg C plus effective Measurement Range 0-360 Deg C Resolution 1Deg Compass Bearing Measurement Range090 Resolution. Display Custom LCD Power Supply 2AA SCADA Installation and project level work of cable laying and converters for existing system configuration shall be extra chargeable.
Other Specifications Download Specification | |
---|---|
Frequency Range | 20Hz - 3 GHz |
Display Type | LCD Backlit (SCADA Opt) |
Capture/Demodulation Bandwidth (Max.) | 25 MHz |
I/Q Memory Depth (256 MB) | Yes |
Trigger | Free Run, Power Trigger, External |
Traces Types | Normal (RMS), Average, Max. Hold, Min. Hold |
Sample Rate | 100 Ms/s |
Ref. Frequency Aging (1 X 10 ^ -7) | 1 X 10 ^ -7 |
Measurement Range (DANL To Max. Safe Input Level) | 20 dBm/Hz - dBm |
Max. Safe Input Level (Should Be Less Than Or Equal To +30 DBm ) | Yes |
Displayed Average Noise Level ( 10 MHz To 3/6/13/26.5 GHz, DANL Should Be Less Than Or Equal To -130 DBm/Hz ) | Yes |
Modulation Analysis | AM, FM, PM, BPSK, QPSK, Pi/4 DQPSK, 8PSK, 16SQAM, 32QAM, 64QAM, 128QAM, 256QAM, GMSK,GFSK |
"Multi Domain Analysis With Stand Alone Software" | Time, Frequency, Modulation, Spectrogram |
Automatic Measurement (Time Mode) | IQ vs Time, Power vs Time, Frequency vs Time, CCDF, crest factor, Pulse Measurements: Pulse Width, Pulse Peak Power, On/Off Ratio, Pulse Ripple, Pulse Repetition Interval, Duty Cycle Pulse-Pulse, Phase, Channel Power, Occupied BW, Frequency Deviation |
Interface | USB, LAN/GPIB |
Operating Temperature | 5 to 40 degree C and 80 % RH at 40 degree C |
Power Requirements (230 +/- 10% VAC, 50 Hz) | Yes |
Rack Mount Kit For Signal Analyzer | Yes |
Availability Of The Test Report From Central Govt./ NABL/ ILAC Accredited Lab As Per Doc Correction Slip No. 38 Dtd 18/1/12 To Be Read In Conjunction With P&C Circular No. 71 Dtd 29/10/14 For Each Model Of Signal Analyzer. | Yes |
TTC No. & Date | CG01A320 |
Name Of The Lab | Ascon Laboratories |
Address Of The Lab | Indore |
Type Test Reports Shall Consist Of Verification Of All The Features & Functional Parameters & Environmental Tests. | Yes |
Dry Heat Test: For 16 Hrs At A Temp. Of 55 Degree C In Accordance With IS: 9000/part-3/section-5/1977 (reaffirmed 2004). The Analyzer Shall Be Checked For All The Functional Requirements Before And After 1-2 Hours Of Recovery Period After The Completion Of The Above Environmental Test. During The Last Half An Hour Of Conditioning Frequency Stability Shall Be Checked. And,the Result Shall Be Satisfactory. | Yes |
Cold Test: For 4 Hrs At A Temp. Of (-)10 Degree C In Accordance With IS: 9000/part-2/section-4/1977(reaffirmed 2004). The Analyzer Shall Be Checked For All The Functional Requirements Before And After 1-2 Hours Of Recovery Period After The Completion Of The Above Environmental Test. During The Last Half An Hour Of Conditioning Frequency Stability Shall Be Checked. And,the Result Shall Be Satisfactory. | Yes |
Damp Heat Test: For 2 Cycles Of 24 Hrs (12 + 12h) Each At A Temp. 40 Degree C & 95% RH In Accordance With IS: 9000/part-5/section-1/1981 (reaffirmed 2004). The Analyzer Shall Be Checked For All The Functional Requirements Before And After 1-2 Hours Of Recovery Period After The Completion Of The Above Environmental Test. During The Last Half An Hour Of Conditioning Frequency Stability Shall Be Checked. And,the Result Shall Be Satisfactory. | Yes |
The Scope Of Supply Shall Include One Set Of Operation And Maintenance Manual For Each Equipment. | Yes |