× All the transactions performed on 2.0 is available here. To access Market click "Home".

Keysight N9912A - 4 GHz Spectrum Analyzer with Cable and Antenna Analyzer (Keysight Ttechnologies)(N9912A with option 104, 230, 303, N9910x-800)

` 1,019,700.00

  • MRP: `1274600.00

Keysight N9912A with 4GHz frequency option combination analyzer offers field ready instrument having dynamic range 49 dB, 42dB directivity, MIL-PRF-28800F Class 2 compliance, Expand the capability with optional Netwok Analyzer, Vector Voltmeter, Interference Analyzer, Power Measurement, 4 Hours battery backup, standard 3-years warranty, Spectrum Analyzer Frequency Range from 100KHz to 4GHz usable to 5KHz. The offered configuration with Standard Cable and Analyzer to measure Return Loss, VSWR, Cable Loss, Smith Chart, 1-port phase, Spectrum Analyzer, Distance to Fault. Refer Keysight datasheet for detailed specifications. Prices indicated including GST @18 percent. Delivery Period 60 Days..

  • Availability: 1 In Stock
  • Min. Qty. Per Consignee: 1
  • Frequency Range - 9 kHz to 3 GHz
  • Aging rate of spectrum analyzer frequency: Better than 1x10-6 /Year - Yes
  • Frequency counter Resolution(min.) - 1 Hz
  • Spectral purity:SSB Phase Noise @10KHz offset shall be less than -90dBc/Hz at 1GHz - Yes
  • Sweep time @span = 0 Hz - 100 micro S to 20 S
Other Specifications Download Specification
Description Of The Item Spectrum analyzers 3GHz
Frequency Range 9 kHz to 3 GHz
Aging Rate Of Spectrum Analyzer Frequency: Better Than 1x10-6 /Year Yes
Frequency Counter Resolution(min.) 1 Hz
Frequency Span 0Hz(zero span),1kHz to 3GHz
Spectral Purity:SSB Phase Noise @10KHz Offset Shall Be Less Than -90dBc/Hz At 1GHz Yes
Spectral Purity :Harmonics @ -40 DBm Level Shall Be Less Than Or Equal To -60dBc Yes
Sweep Time @span >or= 1 KHz 0.1S to 1000 s
Sweep Time @span = 0 Hz 100 micro S to 20 S
Resolution Bandwidth (-3dB) 100 Hz to 3 MHz in steps
Video Bandwidth 30 Hz to 1 MHz in steps
Amplitude : Max Input Level For Protection +30 dBm
Amplitude : Max Input Level For Measurement +20 dBm
Amplitude :level Measurement Uncertainity
Average Noise Level Displayed 300HzRBW,10Hz VBW:
Markers Normal and delta
Trigger Free run, video, External
Display TFT Colour display with provision for three traces
Interface USB and LAN
Measurement Features CN ratio,AM,time domain power measurement,channel power measurement,Phase noise measurement,TOI,OBW
Software For Transfer Of Screen Shot/measurement Data To PC Yes
2-port Vector Network Analyzer Functionality ,for General Purpose Testing Applications In The Field,specially For Cable And Antenna Measurements, Is Available. NA
Frequency Range For 2port Vector Network Analyzer 2 MHz to 3 GHz
Frequency Resolution (CW ON) For 2port Vector Network Analyzer NA Hz
Return Loss For 2port Vector Network Analyzer 0.00 to 60 dB
VSWR For 2port Vector Network Analyzer 1.01 to 10.00
Cable Loss For 2port Vector Network Analyzer 0.00 to 30.00 dB
For 2port Vector Network Analyzer: Calibration With Precision Calibration Components-open, Short, Load Should Be Included To Achieve Directivity > 35 DB Yes
For 2port Vector Network Analyzer: Distance-to-fault In Co-axial Cables Shall Be Possible Yes
Phase Range For 2port Vector Network Analyzer NA degree
Phase Resolution For 2port Vector Network Analyzer NA deg
RF Output Port For 2port Vector Network Analyzer Type N, female, 50Ohm
Dynamic Range For 2port Vector Network Analyzer 60 dB
Nominal Port Power Output For 2port Vector Network Analyzer:Hi Power/Lo-Power Selectable(power Levels Corresponding To Hi And Lo Levels Shall Be Indicated By Bidder. NA
Measurement Of S11 & S21 And Smith Chart Measurement Shall Be Possible NA
For 2port Vector Network Analyzer :Vector Voltmeter For Comparision And Verification Of The Phase Of The Phase-matched Cables Shall Be Available NA
For 2port Vector Network Analyzer:Group Delay Measurement Shall Be Possible.Details To Be Declared NA
Operating Temperature 5 to 40 degree C & 80 RH 40 degree C
Power Requirement 230 +- 10 VAC, 50Hz & External DC Input +12.5V to 15V with rechargeable battery of 1500 mAH or higher
Warranty 3 years
OEM Shall Have ISO:14001 Certification For Their Manufacturing Works yes
One Set Of Operation And Maintenance Manual Shall Be Supplied yes
Standard Accessories Shall Be Supplied yes
Availability Of Test Reports From Central Govt/NABL/ILAC Accredited Lab Covering All The Features & Functional Requirements Including Environmental Tests As Per Specifications yes
Test Report No NTH(SR)/EL(C)/2014/00465I
Test Report Date 28/11/2014
Name Of Lab National Test House
Address Of The Lab NTH, Taramani, Chennai

Price Dekho Price Dekho Price List in india