Synergy HTX Multi-Mode Microplate Reader is a compact, affordable system for 6- to 384-well microplates and Take3 Micro-Volume Plates. Its unique dual optics design provides superior performance for UV-Vis absorbance, fluorescence, luminescence and AlphaScreen/ AlphaLISA workflows. Incubation and shaking plus a dual reagent injector module to meet all your laboratorys assay requirements now and in the future. Synergy HTX is controlled by the easy-to-use, yet powerful, Gen5 Software for data collection, analysis, exporting and reporting. ABSORBANCELight sourceXenon flashDetectorphotodiodeWavelength selectionmonochromatorWavelength range200 - 999 nm, 1 nm incrementsMonochromator bandwidth2.4 nmDynamic range0 - 4.0 ODResolution0.0001 ODPathlength correctionyesMonochromator wavelength accuracy2 nm Monochromator wavelength repeatability0.2 nmOD accuracy1 at 2.0 OD3 at 3.0 ODOD linearity1 from 0 to 3.0 ODOD repeatability0.5 at 2.0 ODStray light.FLUORESCENCE INTENSITYLight sourceTungsten h.

  • Availability: 2 In Stock
  • Min. Qty. Per Consignee: 1
  • Display - High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC
  • Monochromator - Concave holographic grating with 1200 lines/mm
  • Wavelength Range (user selectable) - 340-960 nm
  • Wavelength Accuracy - +-1 nm
  • Type of Detector - Photodiode
Other Specifications Download Specification
Spectrophotometer Type Microprocessor Based
Display High Resolution Touch Screen LCD Colour Graphical display with Built-in-Windows PC
Monochromator Concave holographic grating with 1200 lines/mm
Wavelength Range (user Selectable) 340-960 nm
Wavelength Resolution 0.2 nm
Wavelength Accuracy +-1 nm
Repeatability +-1 nm
Spectral Bandwidth < 2
Type Of Detector Photodiode
Measuring Modes Transmittance (T), Absorbance (ABS), Concentration (CONC) by K Factor and Multi standards.
Operating Mode Single wavelength: Measuring 1, ABS and CONC
Light Source Tungstan Halogen Lamp
Power 230 V +- 10 50Hz
Standard Accessories To Be Supplied 10 nm path length four (4 nos.) Cuvettes matched with 3 percent
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " EN601010
Power Requirements 230 V +- 10, 50 Hz
Operating Temperature 35 Degree C
Operating Humidity (RH) 90 %
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" EN61236
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) ce
Test Report No. MD19.2126
Test Report Date Jun 02, 2015
Name Of The Lab Biotek Winooski, VT 05404 USA
Warranty 1 Year