Microprocessor Flame Photometer Graphical Display Four Element measurement in a single aspiration Na, K, Li, Ca and Ba. Calibration curve fit routines first order and second order. Data displayed in concentration units i.e. ppm meq, automatic filter selection auto calibration feature. Printer attachment, PC connecting compatibility, Sample data storage more than 500 results. Alphanumeric key board for easy data entry. Graphical LCD display thus easy to read, Curve calibration using upto 5 standard, Multiple calibration curve saving facility, Stainless steel burner. Auto Gas Cut off in case of power failure. Provided with 2 Filters Compressor...
Other Specifications Download Specification | |
---|---|
System Is Capable To Measure Alkali And Alkaline Earth Metals (Sodium, Potassium, Calcium, Lithium) In The Samples | Yes |
Ignition System | Auto ignition |
Analyzed Elements | Na+, Ca++, K+, Li+ |
Accuracy | Less than +- 2 % |
Reproducibility | Less than +- 2 % |
Sensitivity | Na+ = 1 ppm, Ca++ = 15 ppm, K+ = 1 ppm, Li+ = 1 ppm |
Display | at least 3 lines LCD |
Detector | Photo conductive cell |
Filters | Sodium, Potassium, Calcium, Lithium |
Aspiration Rate | 3 to 6 ml/min |
Memory | To store and recall 800 samples |
Type Of Printer Interface For Printing Of Samples | NA |
Accessories | Compressor with built in air regulator and air filter |
"Availability Of UL/CE Certification As Per EN 61010-1:2010 Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use. General Requirements " | 9001:2015 |
Power Requirements | 230 V +- 10, 50 Hz AC |
Operating Temperature | 5-50 Degree C Degree C |
Operating Humidity (RH) | 20-50 % |
"Test Reports For EMC/EMI As Per EN 61326-1:2013 Electrical Equipment For Measurement, Control And Laboratory Use. EMC Requirements. General Requirements (Test Reports Are To Be Furnished When Demanded By Buyer)" | 61326-1:2013 |
Availability Of Test Reports From Central Govt / NABL Approved / ILAC Acredited Lab To Prove Conformity To The Specification. (Test Reports Are To Be Furnished When Demanded By Buyer) | Calibration Certificate |
Test Report No. | 1234 |
Test Report Date | 01/01/2018 |
Name Of The Lab | Labtronics |
Warranty | 1 Year |