× All the transactions performed on 2.0 is available here. To access Market click "Home".

ELI 150c electrocardiograph. (mortara)(ELI 150c)

` 144,495.00

  • Offered Price: ` 152100.00
  • MRP: `169000.00

1.Best noise free 10 Second ECG data.2.Unsurpassed acquisition sampling rate of 40 Khz samples/Seconds/Channel for pacemaker detection and 1000 samples/seconds/Channel for storage analysis. 3.Frequency response 0.05 to 300 Hz for Pediatric as well as Adult ECG.4. It is provided with DICOM facility as well RF technology. 5.Alphanumeric KB for easy access and can show the impedance to quality check of connection as well as Rhythm printout. 6.It has 3, 6, 12 Channel formats as well as storage facility of 40 ECG records and up-gradable to 200 ECG records.7. The machine has the feature to connect via computer/LAN/WAN and can transfer the data through USB in PDF Format.8.Display all 12 leads during ECG acquisition as well as lead failure and reversal indication.9.A4 size paper printing roll. 10.Reliable, latest ECG Interpretation software i.e. VERITAS which provide ECG analysis report with reasonable statements 11.Battery or AC operations.12.Up-gradable to wireless module..

  • Availability: 20 In Stock
  • Min. Qty. Per Consignee: 1
  • Functional requirements: electricity (AC-powered) bedside device designed to continuously detect, measure, and display a patient's electrocardiogram (ECG) through leads and sensors attached to the patient, it also typically display heart rate. The device is typically equipped with audible and/or visual alarms that are triggered when the patient's parameters drop below or exceed pre-set limits. Continuous display of patient ECG and heart rate on screen,Allows display of single, 5 lead ECG or simultaneous display of at least 5 waves selected from up to 12 points.Operator can set audiovisual alarm levels for low or high heart rate.Operates from mains voltage or from internal rechargeable battery. - yes
  • Heart rate measurement range to be at least 30 to 250 bpm, with accuracy better than ± 5 bpm - yes
  • Heart rate trend display of at least previous 24 hours - yes
  • Audiovisual alarms required: high and low heart rate (operator variable settings), cardiac arrhythmia, sensor / wire disconnected, low battery - yes
  • weight - 3300 gram
Other Specifications Download Specification
GMDN Name And Code Multichannel Electrocardiographic , CT-1115
Functional Requirements: Electricity (AC-powered) Bedside Device Designed To Continuously Detect, Measure, And Display A Patient's Electrocardiogram (ECG) Through Leads And Sensors Attached To The Patient, It Also Typically Display Heart Rate. The Device Is Typically Equipped With Audible And/or Visual Alarms That Are Triggered When The Patient's Parameters Drop Below Or Exceed Pre-set Limits. Continuous Display Of Patient ECG And Heart Rate On Screen,Allows Display Of Single, 5 Lead ECG Or Simultaneous Display Of At Least 5 Waves Selected From Up To 12 Points.Operator Can Set Audiovisual Alarm Levels For Low Or High Heart Rate.Operates From Mains Voltage Or From Internal Rechargeable Battery. yes
Heart Rate Measurement Range To Be At Least 30 To 250 Bpm, With Accuracy Better Than ± 5 Bpm yes
Heart Rate Trend Display Of At Least Previous 24 Hours yes
Arrhythmia Detection Facility Required; Minimum Gradation Of 1 Bpm yes
Audiovisual Alarms Required: High And Low Heart Rate (operator Variable Settings), Cardiac Arrhythmia, Sensor / Wire Disconnected, Low Battery yes
User's Interface manual
In Built Software And/or Standard Of Communication yes
Weight 3300 gram
Mobility, Portability Protective splashproof case for clean storage and safe transport
Display Must Allow Easy Viewing In All Ambient Light Levels yes
Noise Lavel High-performance baseline filter; AC interference filter 50/60 Hz; low-pass filters 40 Hz, 150 Hz, or 300 Hz dba
Heat Dissipiation: Should Maitain Nominal Temp And The Heat Should Be Disbursed Through A Exhaust Cooling Fan yes
Power Input And Frequency 220 to 240V, 50 Hz
Power Consumption 45W watt
Power Cable Length 3 meter meter
Electrical Protection Provided By Fuses In Both Live And Neutral Supply Lines. yes
"Operating Condition:Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances" yes
"STANDARDS AND SAFETY:FDA / CE Approved Product; Manufacturer / Supplier Should Have ISO 13485 Certificate For Quality Standard. Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (essential Performance Of Electrocardiographs)" yes
Warranty 3 years warranty & one year with min. 3 free servicing. years
Others: List Of Important Spare Parts And Accessories With Their Part Number And Costing. Certificate Of Callibration And Inspection To Be Provided yes

Price Dekho Price Dekho Price List in india