1.Best noise free 10 Second ECG data for storage, analysis and Printing.2.Unsurpassed acquisition sampling rate of 40 Khz samples/Seconds/Channel for pacemaker detection and 1000 samples/seconds/Channel for storage analysis.3.Frequency response 0.05 to 300 Hz for Pediatric as well as Adult ECG acquisition as per American Heart association AHA American College of Cardiology ACC recommendation of low pass setting as 250Hz. 4.It has 3, 6, 12 Channel formats to print and can store up to 20 ECG records in it. 5. Display size -2.9, Remote control keys for easy and fast ECG acquisition or Rhythm printout. 6.Display all 12 leads during ECG acquisition as well as lead failure and reversal indication.7.A4 size paper printing roll. 8.Reliable, accurate latest ECG Interpretation software i.e. VERITAS which provide Adult and Pediatric ECG analysis report with reasonable statements 9.Battery or AC operations.10.Up-gradable to wireless module..
Other Specifications Download Specification | |
---|---|
GMDN Name And Code | Multichannel Electrocardiographic , CT-1115 |
Functional Requirements: Electricity (AC-powered) Bedside Device Designed To Continuously Detect, Measure, And Display A Patient's Electrocardiogram (ECG) Through Leads And Sensors Attached To The Patient, It Also Typically Display Heart Rate. The Device Is Typically Equipped With Audible And/or Visual Alarms That Are Triggered When The Patient's Parameters Drop Below Or Exceed Pre-set Limits. Continuous Display Of Patient ECG And Heart Rate On Screen,Allows Display Of Single, 5 Lead ECG Or Simultaneous Display Of At Least 5 Waves Selected From Up To 12 Points.Operator Can Set Audiovisual Alarm Levels For Low Or High Heart Rate.Operates From Mains Voltage Or From Internal Rechargeable Battery. | yes |
Heart Rate Measurement Range To Be At Least 30 To 250 Bpm, With Accuracy Better Than ± 5 Bpm | yes |
Heart Rate Trend Display Of At Least Previous 24 Hours | yes |
Arrhythmia Detection Facility Required; Minimum Gradation Of 1 Bpm | yes |
Audiovisual Alarms Required: High And Low Heart Rate (operator Variable Settings), Cardiac Arrhythmia, Sensor / Wire Disconnected, Low Battery | yes |
User's Interface | manual |
In Built Software And/or Standard Of Communication | yes |
Weight | 2630 gram |
Mobility, Portability | Protective splashproof case for clean storage and safe transport |
Display Must Allow Easy Viewing In All Ambient Light Levels | yes |
Noise Lavel | High-performance baseline filter AC interference filter 50/60 Hz low-pass filters 40 Hz, 150 Hz, or 300 Hz dba |
Heat Dissipiation: Should Maitain Nominal Temp And The Heat Should Be Disbursed Through A Exhaust Cooling Fan | yes |
Power Input And Frequency | 220 to 240V, 50 Hz |
Power Consumption | 25W watt |
Power Cable Length | 3 meter meter |
Electrical Protection Provided By Fuses In Both Live And Neutral Supply Lines. | yes |
"Operating Condition:Capable Of Operating Continuously In Ambient Temperature Of 0 To 50 Deg C And Relative Humidity Of 15 To 90% In Ideal Circumstances" | yes |
"STANDARDS AND SAFETY:FDA / CE Approved Product; Manufacturer / Supplier Should Have ISO 13485 Certificate For Quality Standard. Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (essential Performance Of Electrocardiographs)" | yes |
Warranty | 3 years warranty & one year with min. 3 free servicing. years |
Others: List Of Important Spare Parts And Accessories With Their Part Number And Costing. Certificate Of Callibration And Inspection To Be Provided | yes |